Local Operation
Operating the Instrument
3
3-55
Table 3-19. Sweep Preferences Fields
ead90f.bmp
Field Preference
Type
[1]
Linear Range, Linear Span, Log Range, Log span
Mode
[2]
Single,
Repetitive
Squelch
Enable, Disable
Trigger Type
[3]
Output, Input, Disable
Trigger Edge
[4]
Rising,
Falling
Prog. Bar Units
%, As Range
[1]
Linear and Logarithmic sweeps – Step size between discrete frequencies is either constant (linear) or logarithmic.
[1]
Range or Span – A Range is presented in terms of a Start and Stop Frequency. Span is an alternative presentation in the form of
Center and Span. The latter tends to be used in very narrow span applications around a center frequency (e.g. Sweep a bandpass
filter). These inputs are transparently converted to Start and Stop values and are bound at this point.
[2]
Single or Repetitive - Like an oscilloscope on single shot or repetitive sweep.
[3]
Enabled or Disabled – When enabled, Squelch is active between all frequency transitions. When disabled, Squelch is active only at
hardware range boundaries.
[4]
Disable, Output or Input. Typically, Trigger is “disabled”. This allows the Sweep to run repetitively or as a single shot when prompted
(Start Sweep key), without accounting for a trigger. Output configures the rear panel BNC to generate a trigger waveform at the start of each
sweep. This trigger waveform can be used to trigger an equivalent sweep in either a Spectrum Analyzer or an Oscilloscope. The action of
sweep remains unaltered. Input configures the rear panel BNC as an input; the Start Sweep is now to “arm” the trigger. The system will now
wait for a trigger at the rear input, before commencing the sweep. This “armed” state is indicated on the Status Bar. Once initiated by a trigger,
the behavior of the Sweep (pausing, stopping, continuing) is un-altered. This feature allows the Instrument sweep to be synchronized with
another instrument.
Note
Trigger Output and Input are Software Trigger features; timing accuracy is typically better than ±1 ms. The trigger
output pulse is delayed by typically 14 to 16 ms from the sweep start to ensure the output signal is settled at the trigger
point.
[5]
Rising or Falling defines the edge polarity generated as a trigger Output or that triggers in the case of Input.
Содержание 9640A Series
Страница 8: ......
Страница 16: ...9640A Instruction Manual viii...
Страница 22: ...9640A Instruction Manual 1 2...
Страница 40: ...9640A Instruction Manual 2 2...
Страница 108: ...9640A Instruction Manual 4A 2...
Страница 138: ...9640A Instruction Manual 4C 2...
Страница 164: ...9640A Instruction Manual 4D 2...
Страница 168: ...9640A Instruction Manual 4D 6...
Страница 170: ...9640A Instruction Manual 4E 2...
Страница 174: ...9640A Instruction Manual 4E 6...
Страница 176: ...9640A Instruction Manual 4F 2...
Страница 184: ...9640A Instruction Manual 4F 10...
Страница 186: ...9640A Instruction Manual 4G 2...
Страница 190: ...9640A Instruction Manual 4G 6...
Страница 192: ...9640A Instruction Manual 5 2...
Страница 272: ...9640A Instruction Manual 5 82...
Страница 274: ...9640A Instruction Manual 6 2...
Страница 284: ...9640A Instruction Manual 7 2...
Страница 309: ...8 1 Chapter 8 Lists of Replaceable Parts Title Page Introduction 8 3 How to Obtain Parts 8 3 Parts Lists 8 4...
Страница 310: ...9640A Instruction Manual 8 2...
Страница 316: ...9640A Instruction Manual 8 8...
Страница 320: ...9640A Instruction Manual A 4...