Fluke 80i-500s Скачать руководство пользователя страница 9

CALIBRATION

If the performance test fails, a qualified technician should
complete the following calibration procedure:

1.  Connect test equipment as shown in Figure 4 (Connection

B) and allow 30 minutes for stabilization. (Do not have the
shunt connected during the calibration procedure.)

2.  Set the 5100B output to 8V at 50 Hz, and enable the 5220A

to operate. This produces 400 ampere-turns in the center
conductor bundle of the toriod coil (8 x 50 turns).

3.  Set the DMM function and range to display 400 mV ac.

4.  Using the Banana-to-BNC Adapter, connect the probe to the

DMM.

5.  Clamp the Current Probe around the coil as shown. (The

probe jaws must be centered around the coil.)

6.  Adjust the calibration potentiometer (accessible next to cable

input to probe handle) to read 403 mV ac on the DMM. (The
adjustment to 403 mV ensures that the probe meets
accuracy specifications where response roll-off occurs.)

7.  Verify calibration by rechecking the test points in Tables 2

and 3.

CAUTION

Remove the shunt before making the last check at
500A. Otherwise, the maximum current rating of the
shunt will be exceeded.

TROUBLESHOOTING

If the 80i-500s does not perform properly, use the following steps
to help isolate the problem:

1.  Inspect the jaw mating surface for cleanliness. If any foreign

material is present, the jaws will not close properly and
errors will result.

2.  Verify that the BNC is connected to an ac volts input on the

measurement instrument. (The Current Probe outputs 1 mV
ac per amp measured.)

3.  Verify that the function selection on the measurement

instrument is correct, i.e., the display vertical resolution is
not to low or high.

REPLACEMENT PARTS

To order replacement parts in the USA, call 1-800-526-4731. To
order outside the USA, contact the nearest service center. Use
the following part numbers:

 

Cable, 600V rms, Safety Insulated

935036

 

Quick Reference Card

936927

 Instruction 

Sheet

936922

Order Fluke Part PM 9053 BNC-to-Banana Adapter to use the
Current Probe with a DMM.

SERIAL NUMBER LOCATION

The serial number date code is located on the handle. The serial
number should be used when ordering parts or requesting
service/calibration information.

Содержание 80i-500s

Страница 1: ...Fluke 80i 500s Manual Get Pricing Availability at ApexWaves com Call Today 1 800 915 6216 Email sales apexwaves com https www apexwaves com probes fluke ac current probes Fluke 2080i 500s...

Страница 2: ...t waveforms and associated harmonics Allows measurement of currents from 1 to 500A ac rms 5 Hz to 10 kHz without breaking into the circuit Provides an output of 1 mV ac per 1 amp ac A passive filter e...

Страница 3: ...are explained in Figure 2 B AC ALTERNATING CURRENT W CAUTION see explanation in manual D Either DC or AC T Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULATION J EARTH Figure 2...

Страница 4: ...500A 5 3 kHz to 10 kHz 1 to 500A not specified 3 db 5 Hz and 10 kHz typical Input Load Impedance of host instrument 1M in parallel with up to 47 pF dV dt max 0 24 mV sec Rise or Fall Time 40 sec W Wor...

Страница 5: ...STRUCTIONS Scope Mode 1 Connect the 80i 500s Current Probe to the desired input channel on the oscilloscope 2 Use a 1 1 probe setting 3 Clamp the current probe around the conductor 4 Press S to select...

Страница 6: ...SE SERVICING INSTRUCTIONS ARE FOR USE BY QUALIFIED PERSONNEL ONLY TO AVOID ELECTRIC SHOCK DO NOT PERFORM ANY SERVICING PROCEDURES UNLESS YOU ARE QUALIFIED TO DO SO READ THE INFORMATION TITLED SAFETY A...

Страница 7: ...check at 500A Otherwise the maximum current rating of the shunt will be exceeded Table 1 Required Test Equipment REQUIRED RECOMMENDED MODEL AC Calibrator Fluke Model 5100B Transconductance Amplifier...

Страница 8: ...ble 2 Performance Test Points 5100B Direct 5100B SETTINGS RMS AMPS MEASURED LOW LIMIT OUTPUT mV HIGH LIMIT OUTPUT mV PHASE SHIFT 0 1A 50 or 60 Hz 5A 4 45 5 55 0 4A 50 or 60 Hz 20A 19 0 21 0 3 Table 3...

Страница 9: ...Remove the shunt before making the last check at 500A Otherwise the maximum current rating of the shunt will be exceeded TROUBLESHOOTING If the 80i 500s does not perform properly use the following ste...

Страница 10: ...ired at the Service Center or at the factory at Fluke s option and returned to purchaser transportation prepaid The instrument should be shipped in the original packaging carton or a rigid container p...

Страница 11: ...on Fluke products call USA 1 888 99 FLUKE 1 888 993 5853 Canada 1 800 36 FLUKE 1 800 363 5853 Europe 31 402 678 200 Japan 81 3 3434 0181 Singapore 65 738 5655 Anywhere in the world 1 425 446 5500 Or v...

Отзывы: