Maximum Conductor Size:

One 30 mm (1.18 in.), 750 kcmils (MCM) THHN, two 25 mm
(0.98 in.), 500 kcmils (MCM) THHN, or one 63 x 5 mm (2.5 x 0.2
in.) Bus Bar.

Output Cable: 1.6 meters (63 in.), BNC termination. Cable and
BNC are rated to 600V ac rms for enhanced safety.

Mechanical Shock: 100G per IEC 68-2-27

INSTRUMENT COMPATIBILITY

The 80i-500s is compatible with any oscilloscope, voltmeter, or
other voltage measuring device that has the following features:

 

Accepts a standard BNC connector. A BNC-to-bannana
adapter (order PM 9053 from Fluke) can also be used with
standard inputs on a digital multimeter (DMM).

 

Range and resolution capable of displaying 1mV of output
per amp of measured current.

 

AC Volts accuracy of 1% or better to take full advantage of
the accuracy of the probe.

 

Input impedance of greater than or equal to 1M

 in parallel

with a maximum of 47 pF.

OPERATING INSTRUCTIONS (Scope Mode)

1.  Connect the 80i-500s Current Probe to the desired input

channel on the oscilloscope.

2.  Use a 1 : 1 probe setting.

3.  Clamp the current probe around the conductor.

4. Press 

S

. to select Scope Mode.

5. Press 

A

.

6. Adjust 

V

 or 

v

 (volts/division) for the best vertical display.

(For line power currents less than 100A, 50 mV/div is a good
first choice.)

7. Adjust 

t

 (time/division) for the best horizontal display.

(For power line frequency, 5 msec/div is a good choice.)

OPERATING INSTRUCTIONS (Meter Mode)

1.  See Figure 3. Connect the 80I-500s Current probe to

Channel A.

2.  Use a 1 : 1 probe setting.

3.  Clamp the current probe around the conductor.

4. Press 

M

to select the Meter Mode.

5. Press 

u

 or 

d

 to select an appropriate meter range.

(The current probe output is 1 mV per ampere measured.)

MEASUREMENT CONSIDERATIONS

Observe the following guidelines for positioning the current probe
jaws:

 

Center the conductor inside the probe jaws.

 

Make sure the probe is perpendicular to the conductor.

 

If possible, avoid measurements close to other current
carrying conductors.

Содержание 80i-500s

Страница 1: ...Fluke 80i 500s Manual Get Pricing Availability at ApexWaves com Call Today 1 800 915 6216 Email sales apexwaves com https www apexwaves com probes fluke ac current probes Fluke 2080i 500s...

Страница 2: ...t waveforms and associated harmonics Allows measurement of currents from 1 to 500A ac rms 5 Hz to 10 kHz without breaking into the circuit Provides an output of 1 mV ac per 1 amp ac A passive filter e...

Страница 3: ...are explained in Figure 2 B AC ALTERNATING CURRENT W CAUTION see explanation in manual D Either DC or AC T Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULATION J EARTH Figure 2...

Страница 4: ...500A 5 3 kHz to 10 kHz 1 to 500A not specified 3 db 5 Hz and 10 kHz typical Input Load Impedance of host instrument 1M in parallel with up to 47 pF dV dt max 0 24 mV sec Rise or Fall Time 40 sec W Wor...

Страница 5: ...STRUCTIONS Scope Mode 1 Connect the 80i 500s Current Probe to the desired input channel on the oscilloscope 2 Use a 1 1 probe setting 3 Clamp the current probe around the conductor 4 Press S to select...

Страница 6: ...SE SERVICING INSTRUCTIONS ARE FOR USE BY QUALIFIED PERSONNEL ONLY TO AVOID ELECTRIC SHOCK DO NOT PERFORM ANY SERVICING PROCEDURES UNLESS YOU ARE QUALIFIED TO DO SO READ THE INFORMATION TITLED SAFETY A...

Страница 7: ...check at 500A Otherwise the maximum current rating of the shunt will be exceeded Table 1 Required Test Equipment REQUIRED RECOMMENDED MODEL AC Calibrator Fluke Model 5100B Transconductance Amplifier...

Страница 8: ...ble 2 Performance Test Points 5100B Direct 5100B SETTINGS RMS AMPS MEASURED LOW LIMIT OUTPUT mV HIGH LIMIT OUTPUT mV PHASE SHIFT 0 1A 50 or 60 Hz 5A 4 45 5 55 0 4A 50 or 60 Hz 20A 19 0 21 0 3 Table 3...

Страница 9: ...Remove the shunt before making the last check at 500A Otherwise the maximum current rating of the shunt will be exceeded TROUBLESHOOTING If the 80i 500s does not perform properly use the following ste...

Страница 10: ...ired at the Service Center or at the factory at Fluke s option and returned to purchaser transportation prepaid The instrument should be shipped in the original packaging carton or a rigid container p...

Страница 11: ...on Fluke products call USA 1 888 99 FLUKE 1 888 993 5853 Canada 1 800 36 FLUKE 1 800 363 5853 Europe 31 402 678 200 Japan 81 3 3434 0181 Singapore 65 738 5655 Anywhere in the world 1 425 446 5500 Or v...

Отзывы: