
Maximum Conductor Size:
One 30 mm (1.18 in.), 750 kcmils (MCM) THHN, two 25 mm
(0.98 in.), 500 kcmils (MCM) THHN, or one 63 x 5 mm (2.5 x 0.2
in.) Bus Bar.
Output Cable: 1.6 meters (63 in.), BNC termination. Cable and
BNC are rated to 600V ac rms for enhanced safety.
Mechanical Shock: 100G per IEC 68-2-27
INSTRUMENT COMPATIBILITY
The 80i-500s is compatible with any oscilloscope, voltmeter, or
other voltage measuring device that has the following features:
•
Accepts a standard BNC connector. A BNC-to-bannana
adapter (order PM 9053 from Fluke) can also be used with
standard inputs on a digital multimeter (DMM).
•
Range and resolution capable of displaying 1mV of output
per amp of measured current.
•
AC Volts accuracy of 1% or better to take full advantage of
the accuracy of the probe.
•
Input impedance of greater than or equal to 1M
Ω
in parallel
with a maximum of 47 pF.
OPERATING INSTRUCTIONS (Scope Mode)
1. Connect the 80i-500s Current Probe to the desired input
channel on the oscilloscope.
2. Use a 1 : 1 probe setting.
3. Clamp the current probe around the conductor.
4. Press
S
. to select Scope Mode.
5. Press
A
.
6. Adjust
V
or
v
(volts/division) for the best vertical display.
(For line power currents less than 100A, 50 mV/div is a good
first choice.)
7. Adjust
t
(time/division) for the best horizontal display.
(For power line frequency, 5 msec/div is a good choice.)
OPERATING INSTRUCTIONS (Meter Mode)
1. See Figure 3. Connect the 80I-500s Current probe to
Channel A.
2. Use a 1 : 1 probe setting.
3. Clamp the current probe around the conductor.
4. Press
M
to select the Meter Mode.
5. Press
u
or
d
to select an appropriate meter range.
(The current probe output is 1 mV per ampere measured.)
MEASUREMENT CONSIDERATIONS
Observe the following guidelines for positioning the current probe
jaws:
•
Center the conductor inside the probe jaws.
•
Make sure the probe is perpendicular to the conductor.
•
If possible, avoid measurements close to other current
carrying conductors.