Output Signal: 1 mV ac/A ac
Influence of Temperature on Accuracy: <0.15% per 10
°
C
(18
°
F) for temperatures from -10 to 18
°
C (14 to 64
°
F) and
from 28 to 50
°
C (82 to 122
°
F).
Basic Accuracy (45 to 65 Hz):
Input Current:
Error:
Phase Shift:
1 to 20A
5% of rdg + 0.3A
Not Specified
20 to 100A
5% of rdg
±3
°
100 to 500A
2% of rdg
±1
°
Extended Accuracy:
For other frequencies, refer to the appropriate input current range
and add the error listed below to the “Basic Accuracy” error.
Input Current:
Error:
(5 to 10 Hz):
1 to 500A
not specified*
(10 to 20 Hz):
1 to 300A
5%
300 to 400A
15%
400 to 500A
25%
(20 to 45 Hz):
1 to 500A
5%
(65 Hz to 3 kHz):
1 to 50A
5%+0.4A
50 to 500A
5%
(3 kHz to 10 kHz):
1 to 500A
not specified*
*-3 db @ 5 Hz and 10 kHz typical
Input Load Impedance (of host instrument): >1M
Ω
in parallel
with up to 47 pF
dV/dt max: 0.24 mV/
µ
sec
Rise or Fall Time: <40
µ
sec
W
Working Voltage: 600V ac rms, in compliance with
IEC-1010-1.
W
Common Mode Voltage: 600V ac rms from earth ground, in
compliance with IEC 1010-1.
Influence of Adjacent Conductor: <5 mA/A
Influence of Conductor Position in Jaw Opening: +/-(1.5% of
r0.1A).
GENERAL SPECIFICATIONS
Dimensions:
66 x 195 x 34 mm (2.6 x 7.7 x 1.3 in.)
Weight:
420g (14.8 oz.)
Temperature:
Operating: -10 to 50
°
C (14 to 122
°
F)
Storage: -40 to 71
°
C (-40 to 160
°
F)
Relative Humidity:
0 to 95% (10 to 30
°
C; 50 to 86
°
F)
0 to 75% (30 to 40
°
C; 86 to 104
°
F)
0 to 45% (40 to 50
°
C; 104 to 122
°
F)
Altitude:
non-operating: 0 to 12000 meters (0
to 39370 ft.) operating: 0 to 2000
meters (0 to 6560 ft.)