PERFORMANCE VERIFICATION

Verify probe accuracy with the test setups shown in Figure 4.
Required test equipment is defined in Table 1. Toriod coil
construction is illustrated in Figure 5. First setup Connection A,
and makes the checks called for in Table 2. (Connect the output
of the probe directly to the signal-in on the phase meter for the
phase check.)

Then disconnect A at the 5100B, and set up Connection B. Make
the first four checks called for in Table 3. (Connect the probe
directly to the phase meter for the phase check.)

CAUTION

Remove the shunt before making the last check at
500A. Otherwise, the maximum current rating of the
shunt will be exceeded.

Table 1. Required Test Equipment

REQUIRED

RECOMMENDED MODEL

AC Calibrator

Fluke Model 5100B

Transconductance Amplifier

Fluke Model 5220A

Digital Multimeter (DMM)

Fluke Model 8050A

Phase Meter

Clarke-Hess Model 6000

0.05

, 2A, non-inductive shunt

Empro Model HA-2-100

Small insulated screwdriver

Spectral

Banana-to-BNC Adapter

Fluke Model PM9053

50-Turn Toriod Coil

(See Figure 5.)

8050A

5220A

50-TURN TOROID COIL

PHASE METER

OVERRANGE

UNDERRANGE

REFERENCE IN

10mV – 300mV

OVERRANGE

UNDERRANGE

REFERENCE IN

10mV – 300mV

MODEL 6000 PHASE METER

DEGREES

REMOTE

POWER

LOCAL

OFFSET

RANGE

ON

+/– 100

0 – 360

5100B DIRECT 

0.05 

0.25W

1%

SHUNT

5100B/5220A

A

5100B

B

Figure 4. Performance Test and Calibration Setup

Содержание 80i-500s

Страница 1: ...Fluke 80i 500s Manual Get Pricing Availability at ApexWaves com Call Today 1 800 915 6216 Email sales apexwaves com https www apexwaves com probes fluke ac current probes Fluke 2080i 500s...

Страница 2: ...t waveforms and associated harmonics Allows measurement of currents from 1 to 500A ac rms 5 Hz to 10 kHz without breaking into the circuit Provides an output of 1 mV ac per 1 amp ac A passive filter e...

Страница 3: ...are explained in Figure 2 B AC ALTERNATING CURRENT W CAUTION see explanation in manual D Either DC or AC T Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULATION J EARTH Figure 2...

Страница 4: ...500A 5 3 kHz to 10 kHz 1 to 500A not specified 3 db 5 Hz and 10 kHz typical Input Load Impedance of host instrument 1M in parallel with up to 47 pF dV dt max 0 24 mV sec Rise or Fall Time 40 sec W Wor...

Страница 5: ...STRUCTIONS Scope Mode 1 Connect the 80i 500s Current Probe to the desired input channel on the oscilloscope 2 Use a 1 1 probe setting 3 Clamp the current probe around the conductor 4 Press S to select...

Страница 6: ...SE SERVICING INSTRUCTIONS ARE FOR USE BY QUALIFIED PERSONNEL ONLY TO AVOID ELECTRIC SHOCK DO NOT PERFORM ANY SERVICING PROCEDURES UNLESS YOU ARE QUALIFIED TO DO SO READ THE INFORMATION TITLED SAFETY A...

Страница 7: ...check at 500A Otherwise the maximum current rating of the shunt will be exceeded Table 1 Required Test Equipment REQUIRED RECOMMENDED MODEL AC Calibrator Fluke Model 5100B Transconductance Amplifier...

Страница 8: ...ble 2 Performance Test Points 5100B Direct 5100B SETTINGS RMS AMPS MEASURED LOW LIMIT OUTPUT mV HIGH LIMIT OUTPUT mV PHASE SHIFT 0 1A 50 or 60 Hz 5A 4 45 5 55 0 4A 50 or 60 Hz 20A 19 0 21 0 3 Table 3...

Страница 9: ...Remove the shunt before making the last check at 500A Otherwise the maximum current rating of the shunt will be exceeded TROUBLESHOOTING If the 80i 500s does not perform properly use the following ste...

Страница 10: ...ired at the Service Center or at the factory at Fluke s option and returned to purchaser transportation prepaid The instrument should be shipped in the original packaging carton or a rigid container p...

Страница 11: ...on Fluke products call USA 1 888 99 FLUKE 1 888 993 5853 Canada 1 800 36 FLUKE 1 800 363 5853 Europe 31 402 678 200 Japan 81 3 3434 0181 Singapore 65 738 5655 Anywhere in the world 1 425 446 5500 Or v...

Отзывы: