Fluke 80i-500s Скачать руководство пользователя страница 6

XXX.X A

3

CHANNEL A

1

X XXX

MANUAL
RANGE

1:1
PROBE
300mV
  

3V

 30V
250V

@RANGE

300A TO 500A

MEASUREMENT

5

Ø1,2,3

N

LOAD

LOAD

2

XXX X

Auto
RANGE

1:1
PROBE

300mV

 

 3V

 30V
250V

@RANGE

1A TO 300A

MEASUREMENT

5

4

M

Figure 3. Setup Guide (Meter Mode)

MAINTENANCE

Before each use, assure continued safety by inspecting the
probe. Look for cracks or missing portions of the probe housing
and output cable insulating cover and for loose or weakened
components. Pay particular attention to the insulation
surrounding the probe jaws. If a probe fails this inspection, tape
is shut to prevent unintended operation. To check probe
performance, complete the “Performance Verification” procedure.

WARNING

THESE SERVICING INSTRUCTIONS ARE FOR USE
BY QUALIFIED PERSONNEL ONLY. TO AVOID
ELECTRIC SHOCK, DO NOT PERFORM ANY
SERVICING PROCEDURES UNLESS YOU ARE
QUALIFIED TO DO SO. READ THE INFORMATION
TITLED “SAFETY” AT THE BEGINNING OF THIS
INSTRUCTION SHEET BEFORE PROCEEDING.

Repairs or servicing not covered in this instruction sheet should
be performed only at a Fluke Service Center. A probe under
warranty will be promptly repaired or replaced (at Fluke’s
discretion) and returned at no charge. See the registration card
for warranty terms. If the warranty has lapsed, the probe will be
repaired or replaced and returned for a fixed fee.

Cleaning and Storage

Periodically wipe the case with a damp cloth and detergent; do
not use abrasives or solvents. Open the jaws and wipe the
magnetic pole pieces with a lightly oiled cloth. Do not allow rust
or corrosion to form on the magnetic core ends.

Содержание 80i-500s

Страница 1: ...Fluke 80i 500s Manual Get Pricing Availability at ApexWaves com Call Today 1 800 915 6216 Email sales apexwaves com https www apexwaves com probes fluke ac current probes Fluke 2080i 500s...

Страница 2: ...t waveforms and associated harmonics Allows measurement of currents from 1 to 500A ac rms 5 Hz to 10 kHz without breaking into the circuit Provides an output of 1 mV ac per 1 amp ac A passive filter e...

Страница 3: ...are explained in Figure 2 B AC ALTERNATING CURRENT W CAUTION see explanation in manual D Either DC or AC T Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULATION J EARTH Figure 2...

Страница 4: ...500A 5 3 kHz to 10 kHz 1 to 500A not specified 3 db 5 Hz and 10 kHz typical Input Load Impedance of host instrument 1M in parallel with up to 47 pF dV dt max 0 24 mV sec Rise or Fall Time 40 sec W Wor...

Страница 5: ...STRUCTIONS Scope Mode 1 Connect the 80i 500s Current Probe to the desired input channel on the oscilloscope 2 Use a 1 1 probe setting 3 Clamp the current probe around the conductor 4 Press S to select...

Страница 6: ...SE SERVICING INSTRUCTIONS ARE FOR USE BY QUALIFIED PERSONNEL ONLY TO AVOID ELECTRIC SHOCK DO NOT PERFORM ANY SERVICING PROCEDURES UNLESS YOU ARE QUALIFIED TO DO SO READ THE INFORMATION TITLED SAFETY A...

Страница 7: ...check at 500A Otherwise the maximum current rating of the shunt will be exceeded Table 1 Required Test Equipment REQUIRED RECOMMENDED MODEL AC Calibrator Fluke Model 5100B Transconductance Amplifier...

Страница 8: ...ble 2 Performance Test Points 5100B Direct 5100B SETTINGS RMS AMPS MEASURED LOW LIMIT OUTPUT mV HIGH LIMIT OUTPUT mV PHASE SHIFT 0 1A 50 or 60 Hz 5A 4 45 5 55 0 4A 50 or 60 Hz 20A 19 0 21 0 3 Table 3...

Страница 9: ...Remove the shunt before making the last check at 500A Otherwise the maximum current rating of the shunt will be exceeded TROUBLESHOOTING If the 80i 500s does not perform properly use the following ste...

Страница 10: ...ired at the Service Center or at the factory at Fluke s option and returned to purchaser transportation prepaid The instrument should be shipped in the original packaging carton or a rigid container p...

Страница 11: ...on Fluke products call USA 1 888 99 FLUKE 1 888 993 5853 Canada 1 800 36 FLUKE 1 800 363 5853 Europe 31 402 678 200 Japan 81 3 3434 0181 Singapore 65 738 5655 Anywhere in the world 1 425 446 5500 Or v...

Отзывы: