Performance Verification Tests
13
Table 3. DC Volts Measurement Verification Points
UUT
Range
Input
DCV
Minimum
1 Year
Maximum
1 Year
Minimum
2 Year
Maximum
2 Year
110 mV
0 mV
-0.017
0.017
-0.017
0.017
110 mV
100 mV
99.959
100.042
99.934
100.067
110 mV
-100 mV
-100.042
-99.959
-100.067
-99.934
1.1 V
0 mV
-0.00006
0.00006
-0.00006
0.00006
1.1 V
1 V
0.99970
1.00031
0.99945
1.00056
1.1 V
-1 V
-1.00031
-0.99970
-1.00056
-0.99945
11 V
0 V
-0.0006
0.0006
-0.0006
0.0006
11 V
10 V
9.9970
10.0031
9.9945
10.0056
11 V
-10 V
-10.0031
-9.9970
-10.0056
-9.9945
110 V
0 mV
-0.006
0.006
-0.006
0.006
110 V
100 V
99.945
100.056
99.895
100.106
110 V
-100 V
-100.056
-99.945
-100.106
-99.895
300 V
0 mV
-0.02
0.02
-0.02
0.02
300 V
300 V
299.84
300.17
299.69
300.32
300 V
-300 V
-300.17
-299.84
-300.32
-299.69