Specifications
9
Loop Power Supply
Setting
1 Year
2 Year
24 Volt
5%
5%
28 Volt
5%
5%
Short circuit protected
Maximum Current: 22 mA
Maximum Input Voltage: 30V dc
Top and Bottom Limits of Ranges with Auto Range On
Range, dc V Measure
Top of Range
Bottom of Range
110 mV
±110.000 mV
0.000 mV
1.1V
±1.10000V
±0.10000V
11V
±11.0000V
±1.0000V
110V
±110.000V
±10.000V
300V
±300.00V
±100.00V
Range, dc V Source
110 mV
+110.000 mV
-10.000 mV
1.1V
+1.10000V
+0.10000V
15V
+15.000V
+1.1000V
Range, ohms Measure and Source
11
Ω
11.000
Ω
0.000
Ω
110
Ω
110.00
Ω
10.00
Ω
1.1 k
Ω
1100.0
Ω
100.0
Ω
11 k
Ω
11.000 k
Ω
1.000 k
Ω
Range, Current Measure
22 mA
+22.000 mA
0.000 mA
110 mA
+110.00 mA
+30.00 mA
Range, Current Source
22 mA
+30.000 mA
0.000 mA
Range, Frequency Measure
100 Hz
109.99 Hz
1.00 Hz
1 kHz
1099.9 Hz
100.00 Hz
10 kHz
10.999 kHz
1.000 kHz
50 kHz
50.00 kHz
10.00 kHz