Performance Verification Tests
19
DC Current Source
Proceed as follows to verify the DC Current Source function.
1.
Connect the UUT to the HP3458A as shown in Figure 7 .
2.
Set the 3458A [NPLC] to 30, and the function to [DC Current].
3.
Set the UUT to DC Current Source (not Simulate Transmitter) function at 2 mA.
4.
Observe that the reading on the 3458A is within the range shown under the
appropriate column in Table 9 .
5.
Continue through the test points, verifying compliance to specifications.
6.
When you are finished, press
c
twice on the UUT to set the Source function to
off. This saves battery life.
TC
mA
mA
V
V
RTD
RTD
SOURCE
300V
MAX
30V
MAX
30V
MAX
30V
MAX
MEAS
CAT
Ω
SENSE
INPUT
HI
LO
GUARD
HI
LO
HP3458
FLUKE 74X
i
gb08f.eps
Figure 7. DC Current Source Verification Connections
Table 9. DC Current Source Verification Points
UUT
Range
UUT
Output
Minimum
1 Year
Maximum
1 Year
Minimum
2 Year
Maximum
2 Year
22 mA
2 mA
1.9965
2.0035
1.9963
2.0037
22 mA
4 mA
3.9963
4.0037
3.9959
4.0041
22 mA
22 mA
21.9945
22.0055
21.9923
22.0077