6200-2
Users Manual
22
Specifications
General Specifications
Size ......................................................................................... 200 mm (L) x 275 mm (W) x 114 mm (H)
Weight ..................................................................................... 3.13 kg
Power Supply .......................................................................... 230 V + 10 % to
−
15 %, 50 Hz ±2 Hz
Power consumption (Tester) ................................................... 13 W typical (idle) 60 W maximum
Operating temperature ............................................................ 0 to +40 °C
Storage Temperature .............................................................. -10 to +60 °C
Relative Humidity .................................................................... non condensing < +10 °C
95 % from +10 to +30 °C
75 % from +30 to +40 °C
Operating Altitude ................................................................... 0 up to 2000 m
Sealing .................................................................................... IP-40 (enclosure), IP-20 (connectors)
EMC ........................................................................................ complies with EN61326-1: Portable
Safety ...................................................................................... Complies with IEC/EN61010-1, CAT II, 300 V,
pollution degree 2
DIN VDE0404-1 and DIN VDE0404-2
IEC/EN 61557 part 1, 2, 4
Test Specifications
The accuracy specification for the display range is defined as ± (%r digit counts) at 23 °C ± 5 °C,
≤
75 % RH.
Between 0 °C and 18 °C and between 28 °C and 40 °C, accuracy specifications may degrade by 0.1 x (accuracy
specification) per °C.
The measurement range meets the service operating errors specified in EN61557-1: 1997, EN61557-2: 1997,
EN61557-4: 1997, DIN VDE0404-2.
Power-on Test
The test indicates reversed L-N, missing PE, and measures the mains voltage and frequency.
Operational Error Measurement Range .................................. 195 V to 253 V
Display Range ......................................................................... 90 V to 264 V
Accuracy at 50 Hz ................................................................... ± (2 % + 3 counts)
Resolution ............................................................................... 0.1 V
Input Impedance ..................................................................... > 1 M
Ω
// 2.2 nF
Maximum Input Mains Voltage................................................ 264 V