5790B
Service Manual
3-38
Table 3-12. DC Voltage Performance Verification Table
Step
No.
Range
Nominal
5790B
Positive
Measurement
5790B
Negative
Measurement
5790B
Average*
Lower Limit
Upper Limit
1
200 mV
60 mV
59.99628 mV
60.00372 mV
2
200 mV
200 mV
199.99110 mV
200.00890 mV
3
700 mV
200 mV
199.99230 mV
200.00770 mV
4
700 mV
600 mV
599.97990 mV
600.02010 mV
5
2.2 V
0.6 V
0.59998680 V
0.60001320 V
6
2.2 V
1.0 V
0.99997800 V
1.00002200 V
7
2.2 V
2.0 V
1.99995600 V
2.00004400 V
8
7 V
2 V
1.999956 V
2.000044 V
9
7 V
6 V
5.999868 V
6.000132 V
10
22 V
6 V
5.999850 V
6.000150 V
11
22 V
20 V
19.999500 V
20.000500 V
12
70 V
20 V
19.99940 V
20.00060 V
13
70 V
60 V
59.99820 V
60.00180 V
14
220 V
60 V
59.99826 V
60.00174 V
15
220 V
200 V
199.99420 V
200.00580 V
16
700 V
200 V
199.9922 V
200.0078 V
17
700 V
600 V
599.9766 V
600.0234 V
18
1100 V
600 V
599.9778 V
600.0222 V
19
1100 V
1049 V
1048.9612 V
1049.0388 V
*The average of the absolute measurement of the Product with normal and reversed polarity stimulus applied. 5790B
Average = (Meas1 + |Meas2|)/2
1. Connect the 5730A Output HI/LO to the Product Input 2 HI/LO using the
low thermal leads. Connect the 5790B Guard to the 5730A Guard. Ensure
the Guard to Ground strap on the 5730A is installed. Set the Product for
Input 2, External Guard and auto range. Allow three minutes for thermal
dissipation after leads are connected.
Содержание 5790B
Страница 6: ...5790B Service Manual iv...
Страница 26: ...5790B Service Manual 1 20...
Страница 104: ...5790B Service Manual 3 66...
Страница 109: ...Maintenance Introduction 4 4 5 hvi026 eps Figure 4 3 Air Filter Access...
Страница 132: ...5790B Service Manual 5 10...
Страница 134: ...5790B Service Manual...
Страница 150: ...5790B Service Manual B 6...