Introduction and Specifications
Electrical Specifications
1
1-11
Absolute AC Non-sine Voltage
Range
[1]
Frequency Range
1-Year, tcal
±
5
°
C,
±
(% of r µV)
2.2 mV
10 Hz - 45 Hz
0.1 + 1.3
45 Hz - 1 kHz
0.1+ 1.3
1 kHz - 20 kHz
0.17+ 1.3
20 kHz - 100 kHz
0.5 + 2.5
7 mV
10 kHz - 45 Hz
0.1 + 1.3
45 Hz - 1 kHz
0.1 + 1.3
1 kHz - 20 kHz
0.17 + 1.3
20 kHz - 100 kHz
0.5 + 2.5
22 mV
10 Hz - 45 Hz
0.1 + 1.3
45 Hz - 1 kHz
0.1 + 1.3
1 kHz - 20 kHz
0.17+ 1.3
20 kHz - 100 kHz
0.5 + 2.5
70 mV
10 Hz - 45 Hz
0.1 + 1.5
45 kHz - 1 kHz
0.1 + 1.5
1 kHz - 20 kHz
0.17 + 1.5
20 kHz - 100 kHz
0.5 + 2.5
220 mV
10 Hz - 45 Hz
0.1 + 1.5
45 Hz - 1 kHz
0.1 + 1.5
1 kHz - 20 kHz
0.17 + 1.5
20 kHz - 100 kHz
0.5 + 2.5
700 mV
10 Hz - 45 Hz
0.1 + 1.5
45 Hz - 1 kHz
0.1 + 1.5
1 kHz - 20 kHz
0.17 + 1.5
20 kHz - 100 kHz
0.5 +
2.5
2.2 V
[2]
10 Hz - 45 Hz
0.1
45 Hz - 1 kHz
0.1
1 kHz - 20 kHz
0.17
20 kHz - 100 kHz
0.5
7 V
10 Hz - 45 Hz
0.1
45 Hz - 1 kHz
0.1
1 kHz - 20 kHz
0.17
20 kHz - 100 kHz
0.5
22 V
[2]
10 Hz - 45 Hz
0.1
45 Hz - 1 kHz
0.1
1 kHz - 20 kHz
0.17
20 kHz - 100 kHz
0.5
70 V
10 Hz - 45 Hz
0.1
45 Hz - 1 kHz
0.1
1 kHz - 20 kHz
0.17
20 kHz - 100 kHz
0.5
[1]
Specifications apply for non-sinusoidal inputs with crest factor <3.0 and with harmonic content band-limited to <1 MHz.
[2]
Crest factor limited to <2.3 for signals greater than 75 % of full scale RMS.
Содержание 5790B
Страница 6: ...5790B Service Manual iv...
Страница 26: ...5790B Service Manual 1 20...
Страница 104: ...5790B Service Manual 3 66...
Страница 109: ...Maintenance Introduction 4 4 5 hvi026 eps Figure 4 3 Air Filter Access...
Страница 132: ...5790B Service Manual 5 10...
Страница 134: ...5790B Service Manual...
Страница 150: ...5790B Service Manual B 6...