C-1
Appendix C
Calibration Constant Information
The constants in these tables are arranged by group. Each group is stored as a
block in nonvolatile memory. The value given for each constant in this list is the
default assigned before the instrument is first calibrated. Defaults are reinstated if
you perform a format of the EEPROM ALL or CAL areas.
Group ZC_BASIC
Name
Default
Function
DAC_Z1
398.0
DAC offset, coarse counts.
DAC_Z2
17500.0
DAC offset, fine counts.
DAC_RATIO
16500.0
DAC coarse/fine count ratio
AD_DIV_Z
0.0
A/D divided (1/6) range offset, in counts
AD_DIV_G
1.397E-8
A/D divided (1/6) range gain, V/count
AD_X1_Z
0.0
A/D x1 range offset, in counts
AD_X1_G
2.328E-9
A/D x1 range gain, V/count
AD_X10_Z
0.0
A/D x10 range offset, in counts
AD_X10_G
-2.328E-10
A/D x10 range gain, V/count
AD_SDL_Z
0.0
A/D SDL range offset, in counts
AD_SDL_G
2.328E-9
A/D SDL range gain, V/count
NULLDAC_Z
0.0
Null DAC offset, in volts
NULLDAC_G
6560.0
Null DAC gain, counts/V
SENSOR_C1
1.0
Sensor linearization
SENSOR_C2
0.0
Sensor linearization
OF_VSQ
0.0
V squared turnover coefficient
REF_CHECK
25.0E-6
A/D - DAC reference difference
Содержание 5790B
Страница 6: ...5790B Service Manual iv...
Страница 26: ...5790B Service Manual 1 20...
Страница 104: ...5790B Service Manual 3 66...
Страница 109: ...Maintenance Introduction 4 4 5 hvi026 eps Figure 4 3 Air Filter Access...
Страница 132: ...5790B Service Manual 5 10...
Страница 134: ...5790B Service Manual...
Страница 150: ...5790B Service Manual B 6...