Appendices
Calibration Constant Information
C
C-21
Group WDC_2_2V
Name
Default
Function
DI_2_2V_WB
3.16228
2_2V range, wideband input, DAC volts per input volt
IA_2_2V_WB
31.6228
2_2V range A/D volts per input volt
Group WAC_2_2V
Name
Default
Function
F1_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 1 (10 Hz)
F2_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 2 (50 Hz)
F3_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 3 (400 Hz)
F4_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 4 (1 kHz)
F5_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 5 (4 kHz)
F6_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 6 (30 kHz)
F7_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 7 (100
kHz)
F8_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 8 (300
kHz)
F9_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 9 (1 MHz)
F10_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 10 (4
MHz)
F11_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 11 (10
MHz)
F12_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 12 (20
MHz)
F13_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 13 (30
MHz)
F14_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 14 (40
MHz)
F15_2_2V_WB
1.0
2_2V range, wideband input, flatness correction 15 (50
MHz)
Group WDC_7V
Name
Default
Function
DI_7V_WB
1.0
7V range, wideband input, DAC volts per input volt
IA_7V_WB
100.0
7V range A/D volts per input volt
Содержание 5790B
Страница 6: ...5790B Service Manual iv...
Страница 26: ...5790B Service Manual 1 20...
Страница 104: ...5790B Service Manual 3 66...
Страница 109: ...Maintenance Introduction 4 4 5 hvi026 eps Figure 4 3 Air Filter Access...
Страница 132: ...5790B Service Manual 5 10...
Страница 134: ...5790B Service Manual...
Страница 150: ...5790B Service Manual B 6...