5790B
Service Manual
3-14
2. Turn on the Product and 5730A/5725A and allow a 30-minute warmup
time.
3. Set the 5730A to EXT SENSE. Verify that the shorting strap is connected
between GUARD and GROUND. Set the Product to EXT GUARD.
4. Select INPUT 1 so that its indicator ring is lit up, then select
Setup
Menu>Calibration>Adjust DC All Main
. The display changes to indicate
first dc calibration step.
5. Set the 5730A to nominal, then use the knob to adjust for the error display
you recorded in Table 3-2. The dc voltage characterization is performed
at cardinal points only. The following rules apply when using the
characterization record for this section: for values that are not
characterized and there is only one characterization point for the
applicable 5730A range, use the same error display as in the
characterization record. For calibration adjust point that is between two
characterized points of the 5730A, use linear interpolation. Set the 5730A
to operate. When the U (unsettled) indicator on the 5730A goes out, push
[Continue].
6. At each step you accept the default value by pushing [Continue]. You do
this because you have already applied your correction in the adjusted
5730A setting. The display tells you that the calibration step is in progress
and informs you with a beep when the step is complete.
7. When the Product completes the step, the next dc step which requires -
2 V dc is presented on the display. Change the 5730A setting accordingly
and do the calibration step as in the previous two steps.
Warning
Some steps in the remainder of this procedure require
application of lethal voltages. Use extreme caution to avoid
contact with live conductors. Set the calibrator to standby and
verify that voltage has returned to zero immediately after each
high voltage calibration step is finished.
Note
For the 200 mV points and below, remove the external sense leads
and set the 5730A to INT SENSE.
8. Follow the Product prompts to continue until the dc calibration is done.
Table 3-4 lists the steps in calibration of the main input. The 1000 V range
is the last dc calibration step.
Содержание 5790B
Страница 6: ...5790B Service Manual iv...
Страница 26: ...5790B Service Manual 1 20...
Страница 104: ...5790B Service Manual 3 66...
Страница 109: ...Maintenance Introduction 4 4 5 hvi026 eps Figure 4 3 Air Filter Access...
Страница 132: ...5790B Service Manual 5 10...
Страница 134: ...5790B Service Manual...
Страница 150: ...5790B Service Manual B 6...