52120A
Users Manual
3-4
Power On and *RST Conditions
All parameters are set to their default values when the Product is energized. Table 3-1 is a
list of items that are set when the Product is turned on and when an *RST command is
received through the GPIB interface. Subsequent changes to these parameters are lost
when you turn off the Product.
Table 3-1. Power On and RST Defaults
Item
Power On State
*RST State
Input Terminals
Voltage
Voltage
Range 2A
2A
Output Terminals
High
High
Load Compensation
Off
Off
*ESR? 128
Unchanged
*ESE 255
Unchanged
*STB? 32
Unchanged
*SRE 255
Unchanged
:OPER:EVENt 0
Unchanged
:QUES:EVENt 0
Unchanged
SCPI Status Registers
Figure 3-2 shows the relationships of the questionable status register, the event status
register, and the operation status register of the Product.