Introduction
and
Specifications
Performance Specifications
1
1-21
52120A/COIL 3 KA 25-Turn Coil
Number of turns
.................................................... 25
Minimum internal jaw dimension to clear wires
26 mm (width) x 36 mm (length)
Maximum input current
........................................ 120 A continuous with built-in 12 V fan on
Maximum voltage
.................................................. 4.5 V rms
Accuracy
Input Current
[1]
Frequency
Effective Current
Amp-turns
Coil Accuracy
[2]
±
(% of Amp-turns + % of 52120A range)
% of Amp-turns
% of 52120A Range
0 A to 100 A
DC
0 to 2500
0.7 %
0.7 %
0 A to 120 A
10 Hz to 65 Hz
0 to 3000
0.7 %
0.7 %
0 A to 120 A
65 Hz to 300 Hz
0 to 3000
0.7 %
0.7 %
0 A to 40 A
300 Hz to 1 kHz
0 to 1000
0.7 %
0.7 %
0 A to 12 A
1 kHz to 3 kHz
0 to 300
0.8 %
1.0 %
0 A to 3 A
3 kHz to 6 kHz
0 to 100
1.5 %
1.0 %
0 A to 1 A
6 kHz to 10 kHz
0 to 50
5.0 %
1.0 %
[1]
The inductance and mutual inductance of the 25 turn coil and clamp that is measured causes a frequency dependent compliance
voltage across the coil. The length and configuration of the cables that connect the current to the coil also have an effect.
Maximum input current is 120 A input at approximately 100 Hz. Maximum current input decreases to approximately 0.8 A at
10 kHz.
[2]
Includes coil/clamp interaction.
52120A/COIL 6 KA 50-Turn Coil
Number of turns
.................................................... 50
Minimum flexible probe length
............................ 500 mm
Maximum input current
........................................ 120 A continuous with built-in 12 V fan on
Maximum voltage
.................................................. 4.5 V rms
Accuracy
Input Current
[1]
Frequency
Effective Current
Amp-turns
Coil Accuracy
[2]
±
(% of Amp-turns + % of 52120A range)
% of Amp-turns
% of 52120A Range
0 A to 100 A
DC
0 to 5000
0.7 %
0.7 %
0 A to 120 A
10 Hz to 65 Hz
0 to 6000
0.7 %
0.7 %
0 A to 120 A
65 Hz to 300 Hz
0 to 6000
0.7 %
0.7 %
0 A to 120 A
300 Hz to 1 kHz
0 to 6000
0.7 %
0.7 %
0 A to 120 A
1 kHz to 3 kHz
0 to 3500
0.8 %
1.0 %
0 A to 25 A
3 kHz to 6 kHz
0 to 1250
1.5 %
1.0 %
0 A to 13 A
6 kHz to 10 kHz
0 to 600
5.0 %
1.0 %
[1]
The inductance and mutual inductance of the 50 turn coil causes a frequency dependent compliance voltage across the coil.
Maximum frequency for 120 A input current is approximately 600 Hz. Maximum current input decreases to approximately 13 A at
10 kHz.
[2]
Includes coil/probe interaction.
Note
The specifications for these coils are at 99 % confidence level and are the combined
accuracy of the coil and a 52120A. If the coils are used with other current sources
the calibration uncertainty of the coils alone is 0.65 % (99 % confidence level) from
0 Hz to 10 kHz.
Содержание 52120A
Страница 2: ......
Страница 6: ...52120A Users Manual iv...
Страница 8: ...52120A Users Manual vi...
Страница 10: ...52120A Users Manual viii...
Страница 12: ...52120A Users Manual 1 2...
Страница 32: ...52120A Users Manual 1 22...
Страница 34: ...52120A Operators Manual 2 2...
Страница 52: ...52120A Operators Manual 2 20...
Страница 54: ...52120A Users Manual 3 2...
Страница 66: ...52120A Users Manual 3 14...
Страница 68: ...52120A Users Manual 4 2...
Страница 72: ...52120A Users Manual 5 2...