![Fluke 52120A Скачать руководство пользователя страница 81](http://html1.mh-extra.com/html/fluke/52120a/52120a_user-manual_2309151081.webp)
Calibration
Introduction
5
5-11
Table 5-4. 20 Amp Range Verification Points (cont.)
5720A
Output
8508A
ACV
filter
8508A
Input
A40B
Measured
Error (%)
Product
Specification
(99 %)
Measurement
uncertainty
(99 %)
TUR
0.8 V, 300 Hz
100
Rear
20 A
- Reference
level
2 V, 300 Hz
100
Front
20 A
±
0.0900 %
±
0.0057 %
16:1
0.8 V, 1 kHz
100
Rear
20 A
- Reference
level
2 V, 1 kHz
100
Front
20 A
±
0.1600 %
±
0.0057 %
>
25:1
0.8 V, 3 kHz
100
Rear
20 A
- Reference
level
2 V, 3 kHz
100
Front
20 A
±
0.5000 %
±
0.0068 %
>
25:1
0.8 V, 6 kHz
100
Rear
20 A
- Reference
level
2 V, 6 kHz
100
Front
20 A
±
1.4000 %
±
0.0068 %
>
25:1
0.8 V, 10 kHz
100
Rear
20 A
- Reference
level
2 V, 10 kHz
100
Front
20 A
±
3.6000 %
±
0.0068 %
>
25:1
120 Amp Range Verification
1.
Push
S
on the Product and 5720A.
2.
Connect the 100 A shunt between the output of the Product and the DMM front
input.
3.
Push
D
on the Product.
4.
Set the 5720A output to +0.6 V dc.
5.
Set the DMM to front input.
6.
Set the 5720A output to +0.6 V dc.
7.
Set the DMM ‘m’ store to the 100 A shunt correction factor.
8.
Set the DMM ‘-c’ and ‘÷z’ stores to 0.6.
9.
Turn on the ‘*m’, ‘-c’ and ‘÷z’, and % math functions on the DMM.
10.
Let the measurement become stable for a minimum of one minute and then record the
measured error in Table 5-5.
11.
Set the DMM ‘-c’ and ‘÷z’ stores to -0.6.
12.
Set the 5720A output to -0.6 V dc and record the measured error.
13.
Set the DMM ‘-c’ and ‘÷z’ stores to 0.8.
14.
Turn on the ‘*m’, ‘-c’ and ‘÷z’, and % math functions on the DMM.
15.
Set the 5720A output to the values in Table 5-5 for the +1 V dc and -1 V dc
verification points.
16.
Set the DMM ‘-c’ and ‘÷z’ stores to -0.8 for the -1 volt measurement.
17.
Record the measured errors.
18.
Set the 5720A output to the values in Table 5-5 for all the ac voltage verification
points and record the measured error. Use the transfer method for all ac
measurements.
For each ac measurement in Table 5-5:
Содержание 52120A
Страница 2: ......
Страница 6: ...52120A Users Manual iv...
Страница 8: ...52120A Users Manual vi...
Страница 10: ...52120A Users Manual viii...
Страница 12: ...52120A Users Manual 1 2...
Страница 32: ...52120A Users Manual 1 22...
Страница 34: ...52120A Operators Manual 2 2...
Страница 52: ...52120A Operators Manual 2 20...
Страница 54: ...52120A Users Manual 3 2...
Страница 66: ...52120A Users Manual 3 14...
Страница 68: ...52120A Users Manual 4 2...
Страница 72: ...52120A Users Manual 5 2...