![Fluke 52120A Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/fluke/52120a/52120a_user-manual_2309151026.webp)
52120A
Users Manual
1-16
General Specifications
Line Power
Voltage range ..................................................... 100 V to 240 V
Frequency ........................................................... 47 to 63 Hz
Voltage variations ...............................................
±
10 % about line voltage
Power consumption ............................................ <1500 VA
Transient overvoltage ......................................... Impulse withstand (overvoltage) Category II of IEC 60364-4-443
Dimensions
(HxWxL)
With feet .............................................................. 192 mm x 432 mm x 645 mm (7.6 in x 17.0 in x 25.5 in)
Without feet ......................................................... 178 mm x 432 mm x 645 mm (7.0 in x 17.0 in x 25.5 in)
Weight
.................................................................... 25 kg (54 lb)
Temperature
Operating ............................................................ 5
°
C to 35
°
C (41
°
F to 95
°
F)
Calibration (tcal) .................................................. 16
°
C to 30
°
C (61
°
F to 86
°
F)
Storage ............................................................... 0
°
C to 50
°
C (32
°
F to 122
°
F)
Transit ................................................................. -20
°
C to +60
°
C (-4
°
F to +140
°
F) <100 hours
Warmup Time
........................................................ Twice the time since last warmed up, to a maximum of 1 hour.
Humidity
(non-condensing)
Operating ............................................................ <80 %, 5
°
C to 31
°
C (41
°
F to 88
°
F) ramping linearly down to 50 % at
35
°
C (95
°
F)
Storage ............................................................... <95 %, 0 to 50
°
C (32
°
F to 122
°
F)
Altitude
Operating ............................................................ 2,500 m (8,200 ft) maximum
Non-Operating .................................................... 12,000 m (39,400 ft) maximum
Shock and Vibration
............................................. MIL-PRF-28800F Class 3
Safety
..................................................................... Complies with EN/IEC 61010-1:2010, CAN/CSA C22.2 No. 61010-1-
04, ANSI/UL 61010-1:2004
EMC
........................................................................ Complies with EN 61326-1:2006, CISPR 11 (EN 55011:2004), FCC
rules part 15, sub part B, Class A
Indoor user only
.................................................... Pollution degree 2
Agency Approvals
................................................
P
,
)
Electrical Performance Limits
Voltage compliance developed across inductive loads may prevent range maximum current output being achieved at
higher frequencies. The appropriate maximum frequency (F
max
) for a given load inductance and current is given by:
F
max
=
4.5
I = Current
2
•
π
•
I
•
L L = Total
inductance
The maximum frequency calculated with this equation is only approximate. Series resistance and parallel capacitance
also affect the maximum achievable frequency.
Input common mode rejection
............................. 80 dB @ DC decreasing linearly to 40 dB at 10 kHz
Input Impedance
Voltage input ....................................................... >1 M
Ω
Current input ....................................................... 10
Ω
Maximum output compliance voltage
................. 4.5 V rms (6.4 V pk), 6.4 V dc. 120 A range maximum compliance
voltage decreases from 4.5 V at 1 kHz to about 3 V at 10 kHz
DC Offset
............................................................... Magnetic remanence that follows abrupt changes in output current
level may cause small changes to DC current offset. It is good practice
to correct for offsets in DC measurements and techniques such as DC
reversal measurement will result in best accuracy.
Содержание 52120A
Страница 2: ......
Страница 6: ...52120A Users Manual iv...
Страница 8: ...52120A Users Manual vi...
Страница 10: ...52120A Users Manual viii...
Страница 12: ...52120A Users Manual 1 2...
Страница 32: ...52120A Users Manual 1 22...
Страница 34: ...52120A Operators Manual 2 2...
Страница 52: ...52120A Operators Manual 2 20...
Страница 54: ...52120A Users Manual 3 2...
Страница 66: ...52120A Users Manual 3 14...
Страница 68: ...52120A Users Manual 4 2...
Страница 72: ...52120A Users Manual 5 2...