Environmental Conditions
Although the Product has optimum durability and trouble-free operation, it must be handled with care.
The Product should not be operated in an excessively dusty or dirty environment. Only use the Product
indoors. Minimize vibrations in the calibration environment. Maintenance and cleaning
recommendations can be found in
.
Forced Air and Convection
Since the Product has a large surface area, a major component of the temperature uncertainty is
caused by changes in convection. Because of this, keep the Product away from areas with large
amounts of air flow or drafts.
Quick Start
Setup
Note
The Product will not heat, cool, or control until the
SET PT.
parameter is Enabled.
Place the Product on a flat surface with at least 15 cm (6 in) of free space around the Product.
Overhead clearance is required. Do not place under a cabinet or structure.
Plug the power cord of the Product into a mains outlet of the proper voltage, frequency, and current
capability (see
Specifications and Environmental Conditions
for power details). Confirm that the
nominal voltage corresponds to that indicated on the power entry model at the back of the Product.
Remove the target cover.
Power on the Product
Toggle the switch on the power entry module to turn on the power to the Product. After a brief self-test,
the controller should begin normal operation. The main screen appears within 30 seconds. If the
Product fails to operate, check the power connection. The display shows the target surface
temperature, and waits for user input before further operation.
Electromagnetic Compatibility (EMC)
International................................. IEC 61326-1: Basic Electromagnetic Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio
frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those
directly connected to a low-voltage power supply network that supplies buildings used for
domestic purposes. There may be potential difficulties in ensuring electromagnetic
compatibility in other environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not
provide adequate protection to radio reception in such environments.
Korea (KCC).................... Class A Equipment (Industrial, Broadcasting, & Communication
Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and
the seller or user should take notice of it. This equipment is intended for use in business
environments and not to be used in homes.
USA (FCC) ...................... 47 CFR 15 subpart B. This product is considered an exempt device
per clause 15.10
[1] The target has a nominal emissivity of 0.928, however it is radiometrically calibrated to minimize lated uncertainties.
Table 1. Specifications (cont.)