TEMP DER
The TEMP DER parameter is the derivative time in seconds that the Product's PID controller uses for
control.
IR CAL
The IR CAL (IR CALIBRATION) menu contains the IR target calibration constants: IR CAL 1, IR CAL 2,
and IR CAL 3. Use the arrow buttons to enter the set-point for each calibration point and push
ENTER
to accept the entry. The calibration points should be selected applicable to model with a low, mid-
range, and high set-point.
IR CAL 1
The IR CAL 1 parameter is the offset in °C for the IR target accuracy at 30 °C.
IR CAL 2
The TEMP 2 parameter is the offset in °C for the IR target accuracy at 40 °C.
WAVELENGTH
The WAVELENGTH parameter can be set to the default 8 µm to 14 µm or USER selectable. This is to
remind the user that the nominal emissivity of the Product pertains to the 8
μ
m to 14
μ
m band.
CALDATE
The CALDATE parameter is the calibration date for the IR target. Use the arrow buttons to enter the
calibration date in the format yyyy,mm,dd.
SYSTEM INFO (View Only)
The SYSTEM INFO (SYSTEM INFORMATION) menu displays manufacturer information regarding the
Product.
MODEL
The MODEL parameter displays the model number of the Product.
SERIAL
The SERIAL (SERIAL NUMBER) parameter displays the serial number of the Product.
FW VERSION
The FW VERSION (FIRMWARE VERSION) parameter displays the firmware version used in the
Product.
CAL DATE
The CAL DATE (CALIBRATION DATE) parameter displays the calibration date of the IR target.