Precision Infrared Calibrator
Maintenance
39
UNIT:TEMP[?] [n]
Read or set the display temperature units, where "n" is a character "C" or "F". Default: C
Read Example: UNIT:TEMP? Response: C
Depending on units setting, a C (Celsius) or F (Fahrenheit) is returned.
Maintenance
With proper care the Product requires very little maintenance.
If a hazardous material is spilled on or inside the Product, the user is responsible for taking the
appropriate decontamination steps as outlined by the National Safety Council with respect to the
material.
Before using any cleaning or decontamination method, except those recommended by Fluke
Calibration, check with an Authorized Service Center (see
) to be sure that
the proposed method does not damage the equipment.
If the Product is used in a manner not in accordance with the equipment design, the operation of the
Product may be impaired or safety hazards may arise.
Caution
To prevent damage to the Product:
•
Do not operate the Product in an oily, wet, dirty, or dusty environment.
•
Do not use harsh chemicals to clean the Product, which may damage the paint. Do
not wipe the front plate (target). If the outside of the Product becomes soiled, wipe it
clean with a damp cloth and mild detergent.
•
Always use the target cover when the Product is not in use and during transport.
•
Handle the Product with care.
•
Avoid knocking or dropping the Product.
•
Keep the target surface of the Product clean and clear of any foreign matter.
•
If the mains supply cord becomes damaged, replace it with a cord with the
appropriate gauge wire for the current of the Product. If there are any questions,
contact an Authorized Service Center (see
) for more
information.
Calibrate the Product annually to maintain its performance to published specifications. For more
information, contact an Authorized Service Center (See