True-rms Digital Multimeter
Calibration Adjustment
27
Table 7. Calibration Adjustment Steps for Firmware Version 1.00 (cont.)
Step
Rotary Switch
Position
Source Value
Settling Time
(seconds)
Calibrating Time
[1]
(seconds)
9
mV AC
0.0 V, 0 Hz
10
5
10
VAC
0.0 V, 0 Hz
4
10
11
VAC
5.0000 V, 60 Hz
10
5
12
VAC
5.0000 V, 65 kHz
[2]
5
35
13
VAC
50.000 V, 60 Hz
6
10
14
VAC
50.000 V, 100 kHz
6
120
15
VAC
250.00 V, 60 Hz
8
2
16
VAC
500.00 V , 60 Hz
6
5
17
VAC
1000.0 V, 60 Hz
7
4
18
VDC
0.0 V, 0 Hz
3
30
19
VDC
5.0000 V, 0 Hz
3
5
20
VDC
5.0 V, 60.0 Hz
4
5
21
VDC
5.0 V, 20.0 kHz
3
35
22
VDC
50.000 V, 0 Hz
3
8
23
VDC
50.0 V, 60.0 Hz
5
15
24
VDC
50.0 V, 20.0 kHz
7
115
25
VDC
500.00 V, 0 Hz
3
7
26
VDC
500.0 V, 60.0 Hz
4
5
27
VDC
1000.0 V, 0 Hz
2
5
28
VDC
1000.0 V, 60.0 Hz
4
5
29
mV DC
50.0 mV, 60.0 Hz
4
3
30
mV DC
500.0 mV, 60.0 Hz
5
3
Connect the 5520A in a 4-Wire configuration and select COMP2-Wire
31
Ω
0
Ω
, zcomp wire2
4
20
32
Ω
500
Ω
, zcomp wire2
5
2
33
Ω
5
k
Ω
, zcomp wire2
5
2
34
Ω
50
k
Ω
, zcomp wire2
5
2
Disconnect 5520A 4-Wire configuration and turn COMP off
35
Ω
500
k
Ω
5
5
36
Ω
5.0
M
Ω
5
5
37
Ω
50.0
M
Ω
5
5
38
Ω
500.0
M
Ω
5
2