True-rms Digital Multimeter
Performance Tests
21
Table 4. 189II/AN Performance Test Steps (cont.)
Display Reading
[1]
Step Function Range
Input
Level
Frequency
or Model
Lower
Limit
Upper
Limit
31.
VAC
500.00 V
500 V
10 kHz
497.75
502.25
32.
VAC
1000.0 V
1000 V
10 kHz
993.5
1006.5
33.
DC V
5.0000 V
4 V
3.9988
4.0012
34.
DC V
50.000 V
-40 V
-40.0120
-39.988
35.
DC V
500.00 V
400 V
399.86
400.14
36.
DC V
1000.0 V
600 V
599.6
600.4
37.
DC V. AC/DC 5.0000 V
200 mV
0 Hz
0.1977
0.2023
38.
DC V, AC/DC 5.0000 V
2 V
5 kHz
1.9640
2.0360
39.
ACV, Peak
5.000 V
4 V p-p, Sq.
wave, +1 V
offset
2 kHz
1.863 2.137
Put Meter’s rotary switch to Capacitance function
Remove Test Leads from 5520A and REL offset
Reconnect Test Leads to 5520A
40 Capacitance
10.00
nF
[3]
5
nF
4.90
5.10
41
Diode Test
3.1000 V
3.5 k
Ω
2.0000 3.1000
42
Diode Test
3.1000 V
0
Ω
N/A N/A
Connect 5520A Auxiliary Output to Meter’s mA/
μ
A and COM Input Terminals
43
μ
A AC
500.00
μ
A
500
μ
A 60
Hz
496.80 503.20
44
μ
A AC
500.00
μ
A 500
μ
A 30
kHz
492.85
[4]
507.15
[4]
45
μ
A AC
5000.0
μ
A 5000
μ
A 30
kHz
4928.5
[4]
5071.5
[4]
46
μ
A DC
500.00
μ
A 500
μ
A
499.42 500.58
47
μ
A DC
5000.0
μ
A 5000
μ
A
4996.0 5004.0
48
mA AC
50.000 mA
4 mA
20 Hz
3.940
4.060
49
mA AC
50.000 mA
30 mA
30 kHz
29.375
[4]
30.625
[4]
50
mA AC
400.00 mA
300 mA
30 kHz
284.60
315.40
51
mA AC
400.00 mA
400 mA
60 Hz
397.55
402.45
52
mA DC
50.000 mA
0.1 mA
0.090
0.110
53
mA DC
50.000 mA
50 mA
49.965
50.035
54
mA DC
400.00 mA
400 mA
399.38
400.62