189II/AN
Service Manual
6
AC Voltage Specifications
Accuracy
Function Range Resolution
20 to 45 Hz
45 to 65 Hz
65 Hz to 10
kHz
10 to 20 kHz 20 to 100 kHz
50 mV
[1]
0.001 mV
1.5 % + 60
0.3 % + 25
0.4 % + 25
0.7 % + 40
3.5 % + 40
[3]
AC mV
500 mV
0.01 mV
1.5 % + 60
0.3 % + 25
0.4 % + 25
0.7 % + 40
3.5 % + 40
5 V
[1]
0.0001 V
1.5 % + 60
0.3 % + 25
0.6 % + 25
1.5 % + 40
3.5 % + 40
[3]
50 V
[1]
0.001 V
1.5 % + 60
0.3 % + 25
0.4 % + 25
0.7 % + 40
3.5 % + 40
500 V
[1]
0.01 V
1.5 % + 60
0.3 % + 25
0.4 % + 25
Not Spec’d
Not Spec’d
AC V
1000 V
0.1 V
1.5 % + 60
0.3 % + 25
0.4 % + 25
Not Spec’d
Not Spec’d
-70 to -62 dB
[2]
0.01 dB
3 dB
1.5 dB
2 dB
2 dB
3 dB
-62 to -52 dB
[2]
0.01 dB
1.5 dB
1.0 dB
1 dB
1 dB
2 dB
-52 to -6 dB
[2]
0.01 dB
0.2 dB
0.1 dB
0.1 dB
0.2 dB
0.8 dB
-6 to +34 dB
[2]
0.01 dB
0.2 dB
0.1 dB
0.1 dB
0.2 dB
0.8 dB
dBV
34 to 60 dB
[2]
0.01 dB
0.2 dB
0.1 dB
0.1 dB
Not Spec’d
Not Spec’d
[1]
Below 5 % of range, add 20 counts.
[2] dBm
(600
Ω
) is specified by 2.2 dB to the dBV range values.
[3]
Add 2.5 % above 65 kHz.
See Detailed Specifications introduction for additional information.
AC Current Specifications
Accuracy
Function Range Resolution
20 to 45 Hz
45 to 1 kHz
1 to 20 kHz
20 to 100 kHz
[4]
500
μ
A 0.01
μ
A
1 % + 20
0.6 % + 20
0.6 % + 20
5 % + 40
AC
μ
A
[3]
5000
μ
A 0.1
μ
A
1 % + 5
0.6 % + 5
0.6 % + 10
5 % + 40
50 mA
0.001 mA
1 % + 20
0.6 % + 20
0.6 % + 20
5 % + 40
AC mA
[3]
400 mA
0.01 mA
1 % + 5
0.6 % + 5
1.5 % + 10
5 % + 40
5 A
0.0001 A
1.5 % + 20
0.8 % + 20
3 % + 40
[4]
Not
Spec’d
AC A
[2]
10 A
[1]
0.001 A
1.5 % + 5
0.8 % + 5
3 % + 10
[4]
Not
Spec’d
[1]
10 A range (10 % to 100 % of range).
[2]
20 A for 30 seconds on, 10 minutes off. >10 A not specified.
[3]
400 mA continuous; 550 mA for 2 minutes on, 1 minute off.
[4]
Verified by design and type tests.
See Detailed Specifications introduction for additional information.
DC Voltage Specification
Accuracy
AC over DC, DC over AC, AC + DC
[2]
Function Range Resolution
DC
[2]
20 to 45 Hz
45 Hz to 1
kHz
1 to 20 kHz
20 to 35 kHz
50 mV
[3]
0.001 mV
0.05 % + 20
[4]
1.5 % + 40
5 % + 40
DC mV
500 mV
0.01 mV
0.025 % + 2
[5]
1.5 % + 40
5 % + 40
5 V
0.0001 V
0.025 % + 2
1.5 % + 40
5 % + 40
50 V
0.001 V
0.025 % + 2
1.5 % + 40
5 % + 40
500 V
0.01 V
0.03 % + 2
Not Spec’d
Not Spec’d
DC V
[1]
1000 V
0.1 V
0.03 % + 2
2 % + 80
0.5 % + 80
Not Spec’d
Not Spec’d
[1]
Add 20 counts in dual display ac over dc, dc over ac or ac+dc.
[2]
AC+DC ranges are specified from 2 % to 140 % of range, except 1000 V is specified from 2 % to 100 % of range.
[3]
When using the relative mode (REL
) to compensate for offsets.
[4]
Add 4 counts/10 mV AC in dual display ac over dc, dc over ac or ac + dc
[5]
Add 10 counts/100 mV AC in dual display ac over dc, dc over ac or ac + dc.