True-rms Digital Multimeter
General Specifications
5
General Specifications
Maximum voltage between any Terminal and Earth Ground:
1000 V
W
Fuse Protection for mA or
μ
A inputs
............. 0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part
only
W
Fuse Protection for A input
............................ 11 A, 1000 V FAST Fuse, Fluke specified part only
Battery Type
.......................................................... 6 AA Alkaline batteries, NEDA 15A IEC LR6
Battery Life
............................................................ 100 hours minimum. 200 hours in Logging mode
Temperature
Operating ............................................................ -20
°
C to 55
°
C
Storage ............................................................... -40
°
C to 60
°
C
Relative Humidity
.................................................. 0 % to 90 % (0
°
C to 37
°
C), 0 % to 65 % (37
°
C to 45
°
C), 0 % to 45 %
(45
°
C to 55
°
C)
Altitude
Operating ............................................................ 3,000 m
Storage ............................................................... 10,000 m
Temperature Coefficient
....................................... 0.05 X (specified accuracy) /
°
C (<18
°
C or >28
°
C)
Vibration
................................................................ Random Vibration per MIL-PRF-28800F Class 2
Shock
..................................................................... 1 meter drop per IEC/EN 61010-1 2
nd
Edition
Size (HxWxL)
......................................................... 22.2 cm x 10.2 cm x 6.0 cm (8.75 in x 4.03 in x 2.38 in)
Weight
.................................................................... 28.0 oz (871 g)
Safety Standards
US ANSI.............................................................. Complies with ANSI/ISA 82.02.01 (61010-1) 2004
CSA..................................................................... CAN/CSA-C22.2 No 61010-1-04 to 1000 V Measurement Category III
and 600 V Measurement Category IV, Pollution Degree 2
UL ....................................................................... UL 61010-1:2004
CE European ...................................................... IEC/EN 61010-1 2
nd
Edition Pollution Degree 2
Electromagnetic Compatibility Standards (EMC)
European EMC ................................................... EN61326-1
Australian EMC...................................................
;
N10140
US FCC............................................................... FCC CFR47: Part 15 CLASS A
Certifications
......................................................... UL, CE, CSA,
;
(N10140),
®
Detailed Specifications
Accuracy:
Accuracy is specified for a period of one year after calibration, at 18
°
C to 28
°
C (64
°
F to 82
°
F), with relative humidity to
90 %. Accuracy specifications are given as: ±( [ % of reading ] + [ number of least significant digits ] ). Accuracy
specification assumes ambient temperature is stable at ±1
°
C. For ambient temperature changes of ±5
°
C, rated accuracy
applies after 2 hours.
True-rms:
AC mV, AC V, AC
μ
A, AC mA, and AC A specifications are ac-coupled, true rms, and are specified from 2 % of range to
100 % of range, except 10 A range is specified from 10 % to 100 % of range.
Crest Factor:
Accuracy is specified with AC crest factor
≤
3.0 at full-scale, increasing linearly to 5.0 at half-scale, except the 1000 V
range, where it is 1.5 at full scale, increasing linearly to 3.0 at half-scale and 500 mV and 5000
μ
A, where it is
≤
3.0 at
80 % of full scale, increasing linearly to 5.0 at half-scale. For non-sinusoidal waveforms add ±(0.3 % of range and 0.1 % of
reading).
AC Floor:
When the input leads are shorted together in the ac functions, the Meter may display a residual reading up to 200 counts.
A 200 count residual reading will cause only a 20 count change for readings at 2 % of range. Using REL to offset this
reading may produce a much larger constant error in later measurements.
AC+DC:
AC+DC is defined as
2
2
dc
ac
+
Содержание 189II/AN
Страница 6: ...189II AN Service Manual iv...
Страница 8: ...189II AN Service Manual vi...
Страница 44: ...189II AN Service Manual 34...