189II/AN
Service Manual
10
Hazardous Voltage
To alert you to the presence of a potentially hazardous voltage, when the Meter detects a
voltage
≥
30 V or a voltage overload (
OL
), the
Y
symbol is displayed.
Required Equipment
Required equipment is listed in Table 2. If the recommended models are not available,
equipment with equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.
Table 2. Required Equipment
Equipment
Required Characteristics
Recommended
Model
Calibrator
AC Voltage Range: 0-1000 Vac
Accuracy:
±
0.0875 %
Frequency Range: 20 Hz – 100 kHz
Accuracy:
±
3 %
DC Voltage Range: 0-1000 Vdc
Accuracy:
±
0.0075 %
Current Range: 0-10A
Accuracy: AC mA -
±
0.15 %
AC A -
±
0.21 %
Frequency range: 1 kHz – 30 kHz
DC mA - +0.0175 %
DC A -
±
0.08 %
Frequency Source: 45 Hz-950 kHz
Accuracy:
±
0.0026 %
Amplitude: 600 mV
Accuracy:
±
5 %
Ohms Range: 0
Ω
- 300 M
Ω
Accuracy:
±
0.0135 %
Capacitance Range: 5 nF
Accuracy:
±
0.55 %
Temperature Range: 0 to 1000
°
C
Accuracy: 0.2 %
Fluke 5520A
Thermocouple Adapter
Accessory
K-type Fluke
80AK
K-type Thermocouple wire K-type, mini-plug on both ends
N/A
Serial Cable, Infrared
Cable, Adapter, Fluke-Custom IRDA Optical to USB,2M
Fluke P/N 2166275