1742/1746/1748
Users Manual
58
Intrinsic Uncertainty ±(% of measurement value + % of power range)
Parameter
Influence
Quantity
Direct Input
[1]
iFlex1500-12
iFlex3000-24
iFlex6000-36
i40S-EL
Clamp: 50 mV/500 mV
Rogowski: 15 mV/150 mV
150 A / 1500 A
300 A / 3000 A
600 A / 6000 A
4 A / 40 A
Active Power P
Active Energy E
a
PF
≥
0.99
0.5 % + 0.005 %
1.2 % + 0.005 %
1.2 % + 0.0075 %
1.7 % + 0.0075 %
1.2 % + 0.005 %
0.1
≤
PF <0.99
see Formula 1
see Formula 2
see Formula 3
see Formula 4
see Formula 5
Apparent Power S
Apparent Energy E
ap
0
≤
PF
≤
1
0.5 % + 0.005 %
1.2 % + 0.005 %
1.2 % + 0.0075 %
1.2 % + 0.0075 %
1.2 % + 0.005 %
Reactive Power Q
Reactive Energy E
r
0
≤
PF
≤
1
2.5 % of measured apparent power/energy
Power Factor PF
Displacement
Power Factor
DPF/cos
φ
-
Reading
±
0.025
Additional uncertainty
(% of power high-
range)
V
P-N
>250 V
0.015 %
0.015 %
0.0225 %
0.0225 %
0.015 %
[1]
Only for calibration laboratories
Reference Conditions:
Environmental: 23 °C
±
5 °C, instrument operating for at least 30 minutes, no external electrical/magnetic field, RH <65 %
Input conditions: Cos
Φ
/PF=1, Sinusoidal signal f=50/60 Hz, power supply 120 V/230 V
±
10 %.
Current and power specifications: Input voltage 1ph: 120 V/230 V or 3ph wye/delta: 230 V/400 V
Input current >10 % of current range
Primary conductor of clamps or Rogowski coil in center position
Temperature Coefficient: Add 0.1 x specified accuracy for each degree C above 28 °C or below 18 °C