Power Quality Logger
Electrical Specifications
57
Flicker P
LT
, P
ST
0 to 20
0.01
5 %
[1]
Only for calibration laboratories
[2]
0 °C … 45 °C: Intrinsic accuracy x 2, outside of 0 °C … 45 °C: Intrinsic accuracy x 3
Accuracy at Reference Conditions (cont.)
Parameter
Range
Maximum Resolution
Intrinsic Accuracy at Reference Conditions
(% of R % of Range)
Power/Energy
Parameter
Direct Input
[1]
iFlex1500-12
iFlex3000-24
iFlex6000-36
i40S-EL
Power Range W, VA, var
Clamp: 50 mV/500 mV
Rogowski: 15 mV/150 mV
150 A / 1500 A
300 A / 3000 A
600 A / 6000 A
4 A / 40 A
Clamp: 50 W/500 W
Rogowski: 15 W/150 W
150 kW / 1.5 MW
300 kW / 3 MW
600 kW / 6 MW
4 kW / 40 kW
Max. Resolution W, VA, var
0.1 W
0.01 kW / 0.10 kW
1 kW / 10 kW
1 kW / 10 kW
0.1 W / 1 W
Max. Resolution PF, DPF
0.01
Phase (Voltage to Current)
[1]
±
0.2 °
±
0.28 °
±
1 °
[1]
In the range of 100 V … 500 V; also known as U
din