Vital Signs Simulator
Detailed Specifications
85
USA (FCC) .......................................................... 47 CFR 15 Intentional Radiators: This device complies with part 15 of
the FCC Rules.
Operation is subject to the following two conditions:
(1) This device may not cause harmful interference.
(2) This device must accept any interference received, including interference that may cause undesired operation.
(15.19).
Changes or modifications not expressly approved by Fluke could void the user's authority to operate the
equipment. (15.21)
Detailed Specifications
Normal-Sinus-Rhythm Waveform
ECG Reference
...................................................... The ECG amplitudes specified are for Lead II (calibration), from the
baseline to the peak of the R wave. All other leads are proportional.
Normal Sinus Rhythm
.......................................... 12-lead configuration with independent outputs referenced to right leg
(RL). Output to 10 Universal ECG Jacks, color-coded to AHA and IEC
Standards.
High-Level Output
................................................. 0.5 V/mV
±
5 % of the ECG amplitude setting available on a BNC
connector.
Amplitude
.............................................................. 0.05 mV to 0.5 mV (0.05 mV steps); 0.5 mV to 5.0 mV (0.25 mV
steps). Other leads are proportional to Lead II (reference lead) in
percentage per:
Lead I .................................................................. 70
Lead II ................................................................. 100
Lead III ................................................................ 30
Lead V1 .............................................................. 24
Lead V2 .............................................................. 48
Lead V3 .............................................................. 100
Lead V4 .............................................................. 120
Lead V5 .............................................................. 112
Lead V6 .............................................................. 80
Amplitude Accuracy
.............................................
±
(2 % of s 0.05 mV)
ECG Rate
............................................................... 10 BPM to 360 BPM in 1 BPM steps
Rate Accuracy
.......................................................
±
1 % of setting
ECG Waveform Selection
..................................... Adult (80 ms) or pediatric (40 ms) QRS duration
ST-Segment Elevation
.......................................... Adult mode only. -0.8 mV to +0.8 mV (0.1 mV steps)
Additional steps: +0.05 mV and -0.05 mV
Power-On Default
.................................................. 60 BPM, 1.0 mV, adult QRS and ST-segment elevation of 0 mV
Pacemaker Waveform
Pacer-Pulse
Amplitude
.......................................................... 0 mV (off),
±
2 mV,
±
4 mV,
±
6 mV,
±
8 mV,
±
10 mV,
±
12 mV,
±
14 mV,
±
16 mV,
±
18 mV,
±
20 mV,
±
50 mV,
±
100 mV,
±
200 mV,
±
500 mV, and
±
700 mV for lead II (reference lead)
Accuracy
Reference lead II ............................................
±
(5 % s 0.2 mV)
All other leads .................................................
±
(10 % s 0.4 mV)
Pacer-Pulse Width
................................................ 0.1 ms, 0.2 ms, 0.5 ms, 1.0 ms, and 2.0 ms
±
5 %
Paced Arrhythmias
............................................... Atrial 80 BPM
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
Atrio-Ventricular sequential
Noncapture (one time)
Nonfunction
Power-On Default
.................................................. Amplitude 10 mV, width 1.0 ms, atrial waveform