Vital Signs Simulator
Simulate Cardiac Output
43
Blood temperature (BT) connector
Injectate temperature (IT) thermistor connector
Injectate temperature (IT) 10-turn potentiometer
CI-3 module/cable
2
1
4
3
Pin1
Pin3
Crimp tube
Thermistor end of cable
ASSEMBLY of the GENERAL PURPOSE CONNECTOR
Viewed from
the top of
the C1-3
PIN
1
FUNCTION
INJECTATE THERMISTOR CONNECTION A
Injectate temperature (IT) thermistor connector
INJECTATE THERMISTOR CONNECTION B
2
NO CONNECTION
4
NO CONNECTION
3
Solder wires to pins 1 and 3
eic224.eps
Figure 26. Cardiac Output Injectate Cable Modification
Note
Fluke Biomedical offers optional adapter cables to connect the
Product to specified brands of cardiac output measurement devices.
•
To examine Hewlett Packard Merlin systems, a cardiac output
adapter and a temperature adapter are necessary.
•
A special adapter is necessary for Gould/SpectraMed Models
SP1445 and SP1465 cardiac output devices.
Contact Fluke Biomedical for availability, price, and part number.
For cardiac output simulation, use the supplied CI-3 module/cable to connect the
Product to the DUT. See Figure 27. If necessary, use the general-purpose
connector.