![Fluke Biomedical ProSim 8 Series Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/fluke-biomedical/prosim-8-series/prosim-8-series_user-manual_2310745025.webp)
Vital Signs Simulator
ECG Function
13
To change the ECG waveform:
1. Push
or
to move the highlight to the
Wave Group
value.
2. Push
.
3. Push
or
to move the highlight to a waveform group name. Table 8 is a
list of the wave groups available in the Product.
Table 8. ECG Wave Groups
Wave Group
NSR (Adult)
Adult Normal Sinus Rhythm
NSR (Pediatric)
Child Normal Sinus Rhythm
Performance
Special waveforms for monitor tests. See
ECG
Performance Tests
.
R-Wave Detection
QRS Detection
Tall T Wave Rejection
Supraventricular
Arrhythmia waveforms. See
Simulate Arrhythmias
.
Premature
Ventricular
Conduction
TV Paced
ACLS
4. When the
Waveform Group
value is highlighted, push
.
To change the heart rate:
1. Push
or
to move the highlight on the display to the
Heart Rate
value.
2. Push
.
3. Push
or
to adjust the heart rate. Each push of a key moves the heart
rate one beat in the direction of the key pushed. The Product simulates a
heart rate of 10 to 360 beats/minute (BPM) set in 1 BPM intervals.
4. Push
to set the heart rate and go back to the ECG screen.
To change the axis:
1. Push
or
to move the highlight to the
Axis
value.
2. Push
.
3. Push
or
to highlight
Intermediate
,
Horizontal
, or
Vertical
.
4. Push
.
Содержание ProSim 8 Series
Страница 8: ...ProSim 8 Users Manual iv...
Страница 10: ...ProSim 8 Users Manual vi...