Vital Signs Simulator
Oximeter SpO2 Optical Emitter and Detector
63
To change the transmission value:
1. From the SpO2 screen, push
or
to highlight the
Transmission
value.
2. Push
.
3. Push
or
to highlight
Dark, Thick Finger
,
Medium Finger
,
Light, Thin
Finger
, and
Neonatal Foot
.
4. Push
to set that transmission value and go back to the SpO2 screen.
You can change the simulated heart rate of the Product from the SpO2 screen.
To change the heart rate:
1. Push
or
to highlight the
Heart Rate
value.
2. Push
.
3. Push
or
to adjust the heart rate. Each push of a key moves the heart
rate one beat in the direction of the key pushed. The range is 10 BPM to
300 BPM.
4. Push
to go back to the SpO2 screen.
To change SpO2 respiration:
1. From the SpO2 screen, push the
or
to highlight the
Respiration
value.
2. Push
.
3. Push
or
to highlight
1%, 2%, 3%, 4%, 5%
or
off
.
4. Push
to set the respiration and go back to the SpO2 screen.
You can test SpO2 under different ambient light conditions. To change the
ambient light value:
1. From the SpO2 screen, push the
or
to highlight the
Ambient Light
value.
2. Push
.
3. Push
or
to highlight an ambient light value. Below is the list of ambient
light values in the Product.
Sunlight
Indoor – 4 kHz
Indoor – DC
Indoor – 5 kHz
Indoor – 50 Hz
Indoor – 6 kHz
Indoor – 60 Hz
Indoor – 7 kHz
Indoor – 1 kHz
Indoor – 8 kHz
Indoor – 2 kHz
Indoor – 9 kHz
Indoor – 3 kHz
Indoor – 10 kHz
4. Push
to go back to the SpO2 screen with the new ambient light value.