Vital Signs Simulator
ECG Function
25
To change the width of the R Wave:
1. Push
or
to highlight the
Width
value.
2. Push
.
3. Push
or
to adjust the width. Each push of a key increases or decreases
the width 20 ms for each key push when the value is 20 ms or above and
2 ms when the value is 20 ms or less. The Product simulates an R Wave
width of 8 ms to 200 ms.
4. When the amplitude is set, push
to enter the value and go back to the
R Wave Detection screen.
Set QRS Detection Test Values
The QRS Detection wave group supplies a signal that you can adjust the width of
the QT interval. This QRS Detection wave includes the R and S waves with
relative amplitudes as specified in EC-13. The R wave is 0.875 of the waveform
amplitude and the S wave is negative at 0.125 of the waveform amplitude. The R
wave up slope is 0.4375 of the waveform width. The R wave down slope is 0.5 of
the waveform width. The S wave up slope is 0.0625 of the waveform width.
To generate a signal for an QRS detection test:
1. Push
to show the ECG screen on the display.
2. Push
or
to highlight the
Wave Group
value.
3. Push
.
4. Push
or
to highlight the
QRS Detection
value in the wave group list.
5. Push
to show the ECG: QRS Detection screen in Figure 10 on the
display.
glh021.bmp
Figure 10. QRS Detection Screen
You can change the rate, amplitude, and width of the QRS wave.
To change the rate:
1. Push
or
to highlight the
Rate
value.
2. Push
.
3. Push
or
to highlight
30
,
60
,
90
,
120
,
200
, or
250 bpm
in the list.
4. Push
to set the rate and go back to the ECG: QRS Detection screen.