190M Series III
Users Manual
8
Figure 4. Scope Connections
XW
Warning
To prevent electrical shock, use the insulation sleeve when you connect the
VPS410 Probe Set without the hook clip or the ground spring.
Probe Type Setup
To obtain correct measurement results the test tool probe type settings must correspond to the
connected probe types.
To select the input A probe setting:
1. Press
A
to show the INPUT A key labels.
2. Press
3
to open the PROBE ON A menu.
3. Use the cursor and
E
to select the probe type Voltage, Current, or Temp.
a. Voltage: select the voltage probe attenuation factor.
b. Current and Temp: select the current probe or temperature probe sensitivity.