Medical ScopeMeter
Replay, Zoom and Cursors
31
6. Use the cursor and
E
to select one of the
Conditions:
and accept the selection.
For external triggering (190M-2-III) continue with:
7. Use the cursor and
E
to select the trigger slope (
Slope
) and level (
Level
).
8. Use the cursor and
E
to select the 0.12V or 1.2V trigger level and accept all recorder
options.
9. Apply a trigger signal to the red and black ext. trigger banana inputs.
During recording samples are continuously saved in deep memory. The last twelve recorded
divisions show on the screen. See
. Use
View All
to display the full memory contents.
Note
To learn more about the Single Shot trigger function, see
Figure 19. Triggered Single Sweep Recording
TrendPlot or Scope Record Analysis
From a TrendPlot or Scope Record you can use the analysis functions CURSORS and ZOOM to
perform detailed waveform analysis. See
.
Replay, Zoom and Cursors
This section covers the capabilities of the analysis functions Cursor, Zoom, and Replay. These
functions can be used with one or more of the primary functions Scope, TrendPlot or Scope
Record. It is possible to combine two or three analysis functions.
A typical application uses these functions:
•
replay the last screens to find the screen of special interest
•
zoom in on the signal event
•
make measurements with the cursors.