Set the sensor supply
1. Make the F113-A safe. If applicable, mind the battery power.
2. Open the F113-A and carefully remove the cable-connectors and the protective cover.
3. Find and set the switches and select the V
out
as required.
4. Close the protective cover and install the cable connectors.
5. Close the F113-A .
Risk of electrocution - High voltage!
Make sure, all the leads to the terminals are disconnected from the F113-A and NEVER
connect the mains power supply to the unit when the protection cover has been removed!
Type PD
Power supply in:
8-24V AC / 10-30V DC
Switch location (typical)
Sensor
V
out
selection
Sensor supply out
A
1
2
3
4
NOTE: Use an AC
autotransformer (spartrafo) with
galvanic isolation.
int
-
off
off
Not suitable for analog sensors
3V DC; <1mA
ext
-
on
on
off
on
off
off
8.2V DC @8V
in
AC / 10V
in
DC
12V DC @10V
in
AC / 14V
in
DC
24V DC @18V
in
AC / 26V
in
DC
Type PF
Power supply in
:
15-24V AC / 20-30V DC
Switch location (typical)
Sensor
V
out
selection
Sensor supply out
A
1
2
3
4
int
-
off
off
Not suitable for analog sensors
3V DC; <1mA
ext
-
on
on
off
on
off
off
8.2V DC @8V
in
AC / 10V
in
DC
12V DC @10V
in
AC / 14V
in
DC
24V DC @18V
in
AC / 26V
in
DC
Type PM
Power supply in
: 115V - 230V AC
Switch location (typical)
Sensor
V
out
selection
Sensor supply out
A
1
2
3
4
int
-
off
off
Not suitable for analog sensors
3V DC; <1mA
ext
-
on
on
off
on
off
off
8.2V DC
12V DC
24V DC
Fig. 9: Sensor supply voltage - Switch setting
Terminal 03-04; alarm / pulse output R2:
This output is designed as a fast output with a maximum frequency of 500Hz. For the use with a
mechanical relay and a continuous signal (pulse) it is recommended to use a maximum frequency of
0.5Hz to prevent an early end-of-life of the mechanical relay.
Terminal 05-06; alarm / pulse output R1:
This output is designed as a slow output with a maximum frequency of 500Hz. For the use with a
mechanical relay and a continuous signal (pulse) it is recommended to use a maximum frequency of
0.5Hz to prevent an early end-of-life of the mechanical relay.