Re-assembling the unit
1. If applicable: check that the battery is installed
correctly in the supply module BSM/RSM and
that the battery cable is correctly inserted into
its connector. See
for battery replacement.
2. If applicable: plug the cable connectors with
connected wiring into the RSM.
3. Position the MEM in front of the housing and
plug the cable connectors with connected
wiring into the back of the MEM.
4. Connect the flat cable between the Supply
Module and the MEM and lock it in place.
5. Making sure the wiring is not damaged or
pinched, insert the MEM into the housing in
one of the four positions (0°; 90°; 180°; 270°
so the display is best readable).
Closing the cover
1. Apply a very thin layer of the specified anti-
seize compound (see
threads and O-ring of the housing.
2. Hold the cover in the correct position for
installation.
3. By hand, screw the cover clockwise onto the
housing until the O-ring is tight to meet the
required IP or TYPE protection rating.
4. Lock the cover in place with the fixing screw.
1,5mm
6.2.4
TEST UNIT
After powering up: check all functions of the display to ensure that all connections are correct.
Action
Result
Notice
Check the battery level indicator
is not lit
If indicator is lit, replace the
battery as described in
Section 7.3: Replace battery [»52]
Only replace with original
batteries supplied by the
manufacturer