2
3
4
5
6
7
Indication
OK:C/CE
CANCEL:
→
Bas Sys Trim
CHK
REVERSE
FORWARD
Bas Sys Trim
CHK
Signal
→
Pulse Output
→
MODBUS
Output
→
Clear Total
Load Settings
SYSTEM SETUP
Bas Sys Trim
CHK
Signal
→
Pulse Output
→
MODBUS
Output
→
Clear Total
Load Settings
Bas Sys Trim
CHK
Low Cutoff %
Max Limit%
Limit Time(S)
Direction
Indication
FORWARD
1
Basic Setup
Bas Sys Trim
CHK
PV Decimal
Total Decimal
Damping(s)
Lcd Rotate
Bas Sys Trim
CHK
Indication
REVERSE
FORWARD
Bas Sys Trim
CHK
Low Cutoff %
Max Limit%
Limit Time(S)
Direction
282.743
Qmax(m
3
/h)
8.) The flow of indication
FORWARD, said flow direction in the same direction with the sign factory settings; REVESRSE, flow
direction in the opposite direction with the factory settings. When the meter on-site installation direction
inconsistent with the direction of the factory (arrow sign on sensor), the instantaneous flow rate is
displayed as "-" .Through the settings to change the flow direction measurement symbols. To change
the sign of the value of flow measurement devices will affect the cumulative values.
Steps
Operation instructions
Interface show
Modify value
Current value
Click “C/CE” two times from
measurement mode into
configuration menu
Click “→” to choice “sys” item
Click “↓” to choice “Signal” item
Click “→” into signal handle menu
Click “↓” to choice Indication item
Click “→” into Indication menu, click
“↓” to setup flow direction
Click “C/CE” quit setup menu ,
LCD show confirm menu
CONVERT PARTS
29
Http: // www.flowmeter.cc
CONVERT