Flomec OM025 Скачать руководство пользователя страница 8

 

 

 

 

 

 

 

 

       Maintenance     13 

 

4.4     Inspection 

(refer Exploded View)

   

Inspect O-rings for damage, chemical attack, deformity or any form of 

deterioration. 

 

 
Remove, inspect and clean the rotors, also check that the magnets have not been chemically attacked. Check the 
measuring chamber for damage or scoring & redress if necessary, the rotor shafts should NOT be loose or able to 
be rotated. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
4.5     Re-assembly of  meter 

(refer Exploded View)

    

When 

r

e-installing the rotors all four magnets MUST be 

visible when both rotors are in place.  Both rotors will only engage correctly if fitted precisely at an orientation of 90 
degrees to each other. Rotate the rotors slowly by hand to ensure they are correctly fitted at the same time check 
the rotor shafts & rotor bearings for wear.

 

 
Fit the O-ring into the groove and assemble the two parts of the meter, the body & cap align with a location pin.  
 
Fit the body cap screws (5) and tighten in a star sequence then torque in the same sequence to 3.5 Nm. This 
sequence and procedure ensures the meter bodies are assembled correctly and evenly. Fit the pulse output board, 
terminal cover or instrument as appropriate. 

14      Fault finding      

 
 

5.0     Fault Finding   

Pulse meters have two distinct sections: the mechanical wetted section housing the rotors 

and the electrical section housing the pulse output board.

 

 
Meters fitted with integral instruments have these two sections plus the instrument. 
The aim of fault finding is to trace the source of the fault to one of these sections. 
If a fault is traced to an instrument section, refer to the relevant instruction manual. 
 
 
Below are basic fault finding steps. Also refer to Trouble Shooting Guide on following page. 
 

Step 1

 - 

Check application, installation and set up.

 

Refer to Mechanical Installation section for installation and application factors that may effect the meter operation 
including pulsation and air entrainment or incorrect meter selection including incorrect flow rate, temperature and 
pressure or materials compatibility.  Refer to Electrical Installation for correct wiring.  
 

Step 2

 - 

Check for blockages.

 

The most common cause of fault/unsatisfactory meter operation, particularly for new or altered installations, is due 
to blockage within the system or meter caused by foreign particles such as weld slag, sealing tape or compound, 
rust, etc. 
 

Step 3

 - 

Ensure flow is present.  

No flow or lower than normal minimum flow may be attributed to a blocked strainer, jammed or damaged rotors 
within the flowmeter,  malfunctioning pump, closed valves or low liquid level in feeder tank. 
 

Step 4

 - 

Ensure oval gears within meter are rotating. 

Rotation of the oval gears can be heard by holding a screw driver blade to the meter body and pressing the handle 
hard against the ear lobe. If necessary test the meter with the flow turned off and turned on to familiarize yourself 
with the audible rotation signature. 
 

Step 5

 - 

Ensure pulses are being generated during flowing conditions. 

A multimeter is often not fast enough to distinguish the pulse train from the reed switch or Hall Effect sensor. An 
oscilloscope will allow you to view the output pulse train. When viewing the Hall effect sensor pulse ensure a pull 
up resistor is installed between the pulse output and the supply voltage (refer electrical installation). 
 

Step 6

 - 

Confirm Instrument Operation. 

If an associated instrument is connected to the flowmeter confirm its operation by simulating a pulse input onto the 
flow input terminals. In most instances a contact closure on the flow input terminals is an adequate simulation.   
 
 
 
 
 
 
 
 
 
 
 
 

Содержание OM025

Страница 1: ...w flomec com au USA 916 Belcher Drive Pelham AL 35124 USA Tel 1 205 378 1050 Fax 1 205 685 3001 Email customerservice trimecus com Website www trimecus com Distributed by Each meter has been calibrated on mineral oil and will contain a small amount of oil residue The oil used is Castrol Diesel Calibration Fluid 4113 product code 055830 Medium capacity positive displacement flowmeters OVAL GEAR SEE...

Страница 2: ...otalisers rate totalisers or batch controllers These instruments also have monitoring and control output options including 4 20mA scaled pulse flowrate alarms and batch control logic preset metering If your flowmeter is fitted or supplied with an instrument please also refer to the relevant instrument instruction manual These flowmeters can be installed within hazardous areas by using the reed swi...

Страница 3: ...en collector phased outputs 4 dc 2 stage batch controller OM040 OM050 accum reset totals pulse output IECEX ATEX approved IECEX ATEX approved IECEX ATEX approved Specifications Model prefix OM015 OM025 OM040 OM050 Nominal size inches 15mm 1 2 25mm 1 40mm 11 2 50mm 2 Flowrange litres min 1 40 10 150 15 250 30 450 Flowrange USGM 0 26 10 6 2 6 40 4 66 8 120 Accuracy 3cp Repeatability Temperature rang...

Страница 4: ...nd air entrapment which can result in erroneous readings or damage the meter on start up Process or safety critical meters should be installed in a by pass section of pipe with isolation valves to enable the meter to be isolated and serviced as required A by pass installation also allows purging of the system during commissioning see Commissioning The meter must be appropriately rated and is typic...

Страница 5: ...stor typically greater than 10K ohms in most instruments which ties the open collector to the available dc voltage level when the Hall sensor is not energized When energized the open collector output is pulled to ground through the emitter 0V 2 3 2 Reed Switch Pulse Output The reed switch output is a two wire normally open SPST voltage free contact ideal for installations without power or for use ...

Страница 6: ...ill have the relevant calibration factor entered into the program of the instrument Please refer to relevant instrument manual for programming details 10 Commissioning Maintenance 3 0 Commissioning Once the meter has been mechanically and electrically installed in accordance with this and any other relevant instrument manual s the meter is ready for commissioning The meter must NOT be run until th...

Страница 7: ...embly with rotor shafts Part No aluminum 1401096 1401094 1401092 stainless steel 1401097 1401093 2 Rotor assembly set aluminum rotors bearings 1424071 1424111 1424099 stainless steel rotors bearings 1424092 1424125 1524011 3 Body O ring size BS153 size BS245 size BS253 viton standard 13031531 13032451 13032531 teflon 13031533 13032453 13032533 4 Meter cap aluminum with M20 conduit entry 1302047 13...

Страница 8: ...tep 1 Check application installation and set up Refer to Mechanical Installation section for installation and application factors that may effect the meter operation including pulsation and air entrainment or incorrect meter selection including incorrect flow rate temperature and pressure or materials compatibility Refer to Electrical Installation for correct wiring Step 2 Check for blockages The ...

Страница 9: ...rminations wires for continuity 1 Rotors fouled 1 Check that rounded teeth are towards base of chamber 2 Check for obstruction due to foreign particles 3 Clean repair or replace rotors 2 1 See instructions for reassembly of meter with particular emphisis on positioning of rotors magnets 3 1 Check terminal connections solder joints 2 Ensure dc voltage is available at Vdc 0v and receiving instrument...

Отзывы: