Set the sensor supply
1. Make the F116-P safe. If applicable, mind the battery power.
2. Open the F116-P and carefully remove the cable-connectors and the protective cover.
3. Find and set the switches and select the V
out
as required.
4. Close the protective cover and install the cable connectors.
5. Close the F116-P.
Risk of electrocution - High voltage!
Make sure, all the leads to the terminals are disconnected from the F116-P and NEVER
connect the mains power supply to the unit when the protection cover has been removed!
Option PD
Power supply in:
8-24V AC / 8-30V DC
Switch location (Typical)
Sensor
V
out
selection
Sensor supply out
A
B
1
2
3
4
NOTE: Use an AC
autotransformer (spartrafo) with
galvanic isolation.
int
int
off
off
Coil 1.2V DC; <1mA
Reed 3V DC; <1mA
ext
ext
on
on
off
on
off
off
8.2V DC; 50mA (max)
12V DC; 50mA (max)
24V DC; 50mA (max)
Option PF
Power supply in
: 24V AC / 24V DC ±10%
Switch location (Typical)
Sensor
V
out
selection
Sensor supply out
A
B
1
2
3
4
int
int
off
off
Coil 1.2V DC; <1mA
Reed 3V DC; <1mA
ext
ext
on
on
off
on
off
off
8.2V DC; 400mA (max)
12V DC; 400mA (max)
24V DC; 400mA (max)
Option PM
Power supply in
: 115V AC - 230V AC ±10%
Switch location (Typical)
Sensor
V
out
selection
Sensor supply out
A
B
1
2
3
4
int
int
off
off
Coil 1.2V DC; <1mA
Reed 3V DC; <1mA
ext
ext
on
on
off
on
off
off
8.2V DC; 400mA (max)
12V DC; 400mA (max)
24V DC; 400mA (max)
Fig. 8: Sensor supply voltage - Switch setting