![Flomec F116-P Скачать руководство пользователя страница 12](http://html.mh-extra.com/html/flomec/f116-p/f116-p_owners-manual_2306730012.webp)
Page 12
FL_F116P_v1702_01_EN
A
COMMUNICATION
C1
speed
1200; 2400; 4800; 9600
C2
address
1 - 247
C3
mode
bus-rtu; bus-asc; off
D
OTHERS
D1
model
F116-P
D2
software version
nn:nn:nn
D3
serial no.
nnnnnnn
D4
password
0000 - 9999
D5
tag-nr
0000000 - 9999999
3.1.2
EXPLANATION OF SETUP-MENU 1 - TOTAL-A
1
TOTAL-A
11 unit
This setting is used to select the engineering unit for the indication of the total (A
and B), the accumulated total (A and B), and the pulse output.
When you change the engineering unit, you must recalculate and
reprogram the K-factor for the (accumulated) total. When you
recalculate and reprogram the K-Factor, the history for
(accumulated) total is not correct anymore, because the
(accumulated) total is not recalculated. For future reference, best
practice is to make a note of the accumulated total before you
program the recalculated K-Factor.
12 decimals
This setting is used to set the amount of digits behind the decimal point for the
(accumulated) total indication (A and B).
13 K-factor
This setting is used to set the K-Factor for the total (A). With the K-Factor, the
flowmeter pulse signals are converted to a quantity. The K-Factor is based on the
number of pulses generated by the flowmeter per selected engineering unit, for
example per m
3
. A more accurate K-Factor (more decimals, as set in decimals
K-Factor) allows for a more accurate operation of the system.
Example 1: Calculating the K-Factor.
The flowmeter generates 2.4813 pulses per liter and the selected unit is m
3
. A
cubic meter consists of 1000 liter which gives 2.4813 pulses*1000 liter=2481.3
pulses per m
3
. So, the K-Factor is 2481.3. Enter for the Flowmeter K-Factor:
24813 and for the flowmeter K-Factor decimals: 1.
Example 2: Calculating the K-Factor.
The flowmeter generates 6.5231 pulses per gallon and the selected engineering
unit is gallons. So, the K-Factor is 6.5231. Enter for the Flowmeter K-Factor:
65231 and for the Flowmeter K-Factor decimals: 4.
When you recalculate and reprogram a new K-Factor, the history for
(accumulated) total is not correct anymore, because the
(accumulated) total is not recalculated. For future reference, best
practice is to make a note of the accumulated total before you
program the recalculated K-Factor.
14 decimals K-factor
This setting is used to set the amount of digits behind the decimal point for the
K-Factor (A).
Содержание F116-P
Страница 2: ...Page 2 FL_F116P_v1702_01_EN ...
Страница 37: ...Page 37 FL_F116P_v1702_01_EN Fig 29 F116 P AP CT OT PX XI Output loop powered IIB IIC IIIC ...
Страница 48: ...Page 48 FL_F116P_v1702_01_EN DECLARATION OF CONFORMITY ...
Страница 52: ...Page 52 ...