The FLIR Exx-EST camera can be used to screen passengers, customers, or employees
for signs of elevated skin temperature.
Screening with the FLIR Exx-EST can be performed in different ways:
• Using one of the built-in Screening modes; Auto, Button, or Operator mode. For more
information, see section 8
Overview — Screening camera modes
.
• Using the camera in combination with the desktop software FLIR Screen-EST. For
more information, refer to the FLIR EST Thermal Screening documentation, see sec-
tion 3.1
.
For accurate and efficient screening, it is important to set up and operate the screening
station correctly. The setup and operation depends on the FLIR EST solution (Auto, But-
ton, or Operator mode, or FLIR Screen-EST) you choose to use.
9.1
Auto and Button mode screening
9.1.1
Set up the screening station
The setup procedure includes the following main steps:
1. Plan and prepare the screening station.
2. Prepare the camera.
3. Prepare the button (Button mode).
4. Place the camera and display.
5. Define the screening position.
6. Adjust the angle of the camera.
7. Test the setup.
8. Place the queueing system and supporting materials.
9. Clearly mark the screening position.
10. Mark the positions of all equipment.
For detailed instructions, see chapter 10
Setup — Screening station (Auto and Button
.
9.1.2
Operate the screening station
The operation of the screening station includes the following operator tasks:
1. Prepare the system for a new screening session.
2. Auto mode: The screening normally does not require any action from the operator.
Depending on the company specific workflow, the operator may need to be available
for guidance and alarms.
3. Button mode: During the screening, you guide the visitors to a good screening and
trigger the screening by pressing a Bluetooth button.
For detailed instructions, see chapter 12
Operation — Operator instructions (Auto and
.
9.2
Operator mode screening
In Operator mode, an operator actively screens people when they are in the correct posi-
tion. Before the screening, the operator must record reference samples.
For detailed instructions, see chapter 13
.
9.3
FLIR Screen-EST screening station
For detailed instructions on how to setup, configure, and operate a screening station with
FLIR Exx-EST and the FLIR Screen-EST application, refer to the FLIR EST Thermal
Screening documentation, see section 3.1
.
#T810583; r. AA/75420/75423; en-US
38