7 Connection
7.1 Transducers
PIOX S502ID
2020-01-24, UMPIOX_S502IDV1-2EN
34
• Fix the extension cable and the external shield to the shield terminal.
• Connect the extension cable to the terminals of the junction box.
Important!
The external shield of the extension cable must not have electrical contact to the junction box. Therefore, the
extension cable has to remain completely insulated up to the shield terminal.
Tab. 7.5:
Terminal assignment
terminal
connection (extension cable)
TV
white or marked cable (core)
TVS
white or marked cable (internal shield)
TRS
brown cable (internal shield)
TR
brown cable (core)
shield terminal
external shield
Fig. 7.7:
Connection of the extension and transducer cable to the junction box
1 – extension cable
2 – external shield
3 – shield terminal
4 – cap nut
5 – compression part
6 – basic part
7 – connection of the extension cable
8 – connection of the transducer cable
cable gland
4
5
6
10
0
m
m
20
m
m
10
m
m
1
2
3
7
8
Содержание PIOX S502ID
Страница 1: ...PIOX S502ID UMPIOX_S502IDV1 2EN Operating instruction ...
Страница 62: ...Annex A Conformity declarations PIOX S502ID 2020 01 24 UMPIOX_S502IDV1 2EN 62 A Conformity declarations ...
Страница 63: ......
Страница 64: ......