13 Settings
13.2 Function test of the outputs
PIOX S502ID
2020-01-24, UMPIOX_S502IDV1-2EN
58
If the multimeter displays the entered value, the output functions correctly.
• Select
yes
to repeat the test or
no
to return to the menu item
Special setups
.
• Press ENTER.
Binary output
• Select the menu item
Check outputs
.
• Press ENTER.
• Select a binary output.
• Press ENTER.
• Select
Opto-Relay OFF
in the scroll list
Output Test
to test the de-energized state of the output.
• Press ENTER.
• Measure the resistance at the output. The value has to be high ohmic.
• Select
yes
.
• Press ENTER.
• Select
Opto-Relay ON
in the scroll list
Output Test
to test the energized state of the output.
• Press ENTER.
• Measure the resistance at the output. The value has to be low ohmic.
• Select
yes
to repeat the test or
no
to return to the menu item
Special setups
.
• Press ENTER.
SPECIALS\Special setups\Check outputs
I1=
10.0 mA
Again no >YES<
Select output
Current B1
(
)
B1:Output Test
Opto-Relay OFF
B1=off
Again no >YES<
B1:Output Test
Opto-Relay ON
B1=off
Again no >YES<