7 Connection
7.1 Transducers
PIOX S502ID
2020-01-24, UMPIOX_S502IDV1-2EN
30
7.1.2
Connection of the extension cable to the transmitter
• The extension cable is connected to the transmitter via the transducer connection.
• Remove the blind plug for the connection of the transducer cable.
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut and the compression part.
• Prepare the extension cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the basic part into the transmitter housing.
• Insert the extension cable into the housing.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the extension cable to the terminals of the transmitter.
Fig. 7.3:
Connection of the transducer cable with plastic jacket and stripped ends to the transmitter
1 – cap nut
2 – compression part
3 – basic part
4 – external shield, brushed back
Notice!
For good electromagnetic compatibility (EMC), it is important to ensure good electrical contact between the external
shield and the cap nut (and thus the housing).
Tab. 7.2:
Terminal assignment
terminal
connection
AV
white or marked cable (core)
AVS
white or marked cable (internal shield)
ARS
brown cable (internal shield)
AR
brown cable (core)
AV
AV
S
AGN
ARS
AR
70
m
m
20
m
m
10
m
m
4
cable gland
1
2
3