21
Extremely dry root canal
A dry canal might disturb the electrical signal.
5.
Troubleshooting
Failure
Possible cause
Solutions
The motor handpiece does not
rotate.
Device in EAL mode
EAL mode is only for canal
measurement.
Changing to CW, CCW, REC or ATR
mode.
There is a continuous beep
sounds after starting the
motor handpiece.
The continuous beep sound is
indicating that the motor
handpiece is under CCW mode.
Stop the motor handpiece and
change the operating mode to CW
Mode.
Contra-angle calibration failure Calibration failure caused by
strong resistance of contra-
angle
Clean the contra-angle and
recalibrate after oil injection.
Motor handpiece heating
Under Reciprocating Motion
Mode, the using time is too
long.
Stop use. Use after the
temperature of motor handpiece
drops.
The time of endurance
becomes shorter after
charging.
Battery capacity becomes
smaller.
Contact local distributor.
No sound
Beeper Volume set to 0.
Vol.0: Mute.
Set Beeper Volume to 1,2,3,4.
The continuously rotating file
is stuck at the root canal.
Incorrect specification setting.
Too high load torque of file.
Choose CCW Mode, start the
motor handpiece, and take the
file out.
6.
Cleaning, Disinfection and Sterilization
6.1
Foreword
For hygiene and sanitary safety purposes, the contra-angle (including the O-ring), the lip hook,
the file clip, the protective silicon cover, and the touch probe must be cleaned, disinfected and
sterilized before each usage to prevent any contamination. This concerns the first use, as well as
all subsequent uses.
6.2
General recommendations
a)
After each use, all the objects that were in contact with infectious agents should be
cleaned using towels impregnated with a disinfection agent.
b)
Use OXYTECH
®
disinfecting solution or any other disinfection agent compliant with local
national regulations (such as VAH/DGHM-listing, CE marking, FDA, and Health Canada
approval) and in accordance with the IFU of the disinfecting solution manufacturer.