4
Power Requirements
Models 825
1800 W
Models 838
3300 W
Models 851
3300 W
Chamber Volumes
Models 825
2.5 cu ft
Models 838
3.8 cu ft
Models 851
5.0 cu ft
Chamber Dimensions
(W x D x H)
Models 825
18 x 18 x 13.5 in
Models 838
18 x 18 x 20 in
Models 851
18 x 18 x 26.5 in
Performance Characteristics
Operating Range
50 to 325°
Average Uniformity @ 200°*
±2°C
Resolution 1°C
Control Sensitivity
±0.25°C
Recovery Time @ 200°C
Model 825
1.5 minutes
Model 838
1.5 minutes
Model 851
3.5 minutes
Average Rise Time to 325°C
Model 825
60 minutes
Model 838
40 minutes
Model 851
40 minutes
Air Exchanges
Model 825
61/hr
Model 838
41/hr
Model 851
31/hr
BTU/hr Output
@100°C
@200°C
Model 825
1085
2650
Model 838
1355
3380
Model 851
1690
3950
*as per ASTM E-145
Electrical Requirements
Model 825
Cat. No. 13-247-825F/FC
120 V, 50/60 Hz
Cat. No. 13-247-826F/FC
240 V, 50/60 Hz
Model 838
Cat. No. 13-247-838F/FC
240 V, 50/60 Hz
Cat. No. 13-247-839F/FC
240 V, 50/60 Hz
Model 851
Cat. No. 13-247-851F/FC
240 V, 50/60 Hz
Cat. No. 13-247-852F/FC
240 V, 50/60 Hz