Replacing the Handle
To replace a defective door handle, perform the steps
below:
1.
Remove the two mounting screws holding latch
cover in place.
2.
Remove the two mounting screws holding
defective handle in place.
3.
Mount the replacement handle using two
screws.
4.
Adjust bottom nut (13/16) from end of shaft.
5.
Secure latch cover in place with two screws.
Adjusting the Door Cam
Due to handling in shipment or to heat distortion with use,
the door cam may require adjustment. To facilitate proper
closing and sealing of door, steps 1 through 6 may have
to be performed concurrently. NOTE: See Warning
above!
To adjust the door cam, perform the following:
1.
Open door and remove screws holding latch
cover in place.
2.
Locate nuts securing tongue on cam shaft.
3.
Loosen but do not remove outside nut.
4.
Adjust inside nut, one full turn clockwise draws
door 1/16” closer to cabinet when door is
closed.
5.
Secure cam tongue in place by tightening out-
side nut.
6.
Secure latch cover in place with two screws.
31
S
ERVICE
Warning
Allow oven to cool to ambient tem-
perature before attempting repair.