To run a program currently in memory, perform the follow-
ing steps:
1.
Press the Run Key once. The Run indicator
light begins to flash and the displays indicate
which program in memory is ready to run. To
change program numbers, press the Up or
Down Arrow Key.
2.
Press the Menu Key when the desired program
number is shown in the display. The display will
next prompt for the user to enter the program
step to begin the program run (usually step #1).
3.
Use the Up or Down Arrow Key to change the
programs starting point. When the desired start-
ing step is displayed, press the Run Key to
begin. The program is running when the Run
LED is lit.
To stop a running program,
press the Run Key. Oven
operation then returns to the Control mode, with the set
temperature at the last value achieved by the program.
To resume a stopped program
, perform the following
steps:
1.
Press the Run Key. The Run LED begins to
flash and the display indicates “1 Prog”.
2.
Press the Menu Key twice. The resume pro-
gram command should appear on the displays.
The Actual Display will show the program num-
ber that was most recently run.
3.
Press the Run Key again. The controller will
continue running the program at the point it was
terminated. The Run LED will again be lit.
While a ramp and soak program is running, the status of
the current program step may be viewed. With the Run
LED lit to indicate the program is running, successively
press the Menu Key. Each control parameter associated
with the current program step then is displayed sequen-
tially. The parameter type is indicated in the Set Display,
while the value for the parameter appears in the Actual
Display. A final press of the Menu Key returns the dis-
plays to tracking the actual and set temperatures.